Default
Module
Item
MDF-DU502VH
MDF-DU702VH
Range
Alarm
High temp. alarm
temp.
+10
℃
+10
℃
+5
℃
~
+40
℃
Low temp. alarm
temp.
-10
℃
-10
℃
-5
℃
~
-40
℃
Temp. alarm delay
time
15 min
15 min
0 min
~
15 min
Door alarm delay
time
2 min
2 min
0 min
~
15min
Ring back function
ON
ON
ON, OFF
Ring back time
30 min
30 min
1 min
~
99min
Remote alarm
linkage
OFF
OFF
ON, OFF
Filter alarm temp.
45
℃
45
℃
※
1
Filter alarm release
temp.
Alarm temp.
-5
℃
or less
Alarm temp.
-5
℃
or less
fix
Comp protection
alarm temp.
50
℃
50
℃
※
1
Comp protection
release alarm
temp.
Ambient temp.
+5
℃
or less
Ambient temp.
+5
℃
or less
fix
Sensors open
alarm
4950mV
4950mV
fix
Sensors short
alarm
25mV
25mV
fix
Battery SW
judgment
100mV
100mV
fix
Port heater voltage
judgement
0.2V
0.2V
fix
Range over of
ambient temp.
0
~
35
℃
0
~
35
℃
fix
Overload running
notice
5 day
5 day
6hour,2day,5day
※
1
Battery change
notice
1022 day
1022 day
fix
Touch panel
communication
error
130 sec
130 sec
fix
Backup battery
changing failure
0.3V / 6 hour
0.3V / 6 hour
fix
Calibration value
Chamber temp.
sensor PCB
calibration value
0
℃
0
℃
※
1
Chamber temp.
sensor calibration
value
0
℃
0
℃
※
1
Ambient temp.
sensor calibration
0
℃
0
℃
※
1
Cascade temp.
sensor calibration
value
0
℃
0
℃
※
1
Filter sensor
calibration value
0
℃
0
℃
※
1
Security operation
Touch panel OFF
4.45V
4.45V
fix
※
1 Variable in service menu
-32-