43

 

 

 

  Automatic N

2

 gas (O

2

 gas) cylinder changeover 

 
It is the function that switching the empty N

2

 gas (O

2

 gas) cylinder to the other one, when the N

2

 gas (O

2

 

gas) cylinder being used becomes empty. 

Note:

 A gas auto changer MCO-21GC which changes the empty CO

2

 gas cylinder into the other one 

automatically, is available as an optional accessory (refer to page 83

86). 

 
Refer to page 23

24 for connecting N

2

 gas cylinder (O

2

 gas cylinder). 

 
N

2

/O

2

 gas supply line indicator A or B being used is lighted in the Top screen. 

 

N

2

/O

2

 gas supply line indicator A and B   

 

N

2

/O

2

 gas supply line select key

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

N

2

/O

2

 gas supply line being used

Summary of Contents for MCO-170M Series

Page 1: ...ons carefully before using this product and save the operating instructions for future use See page 101 for all model numbers MCO 170M MCO 170MUV MCO 170MUVH Series O2 CO2 Incubator MCO 170M MCO 170MUV MCO 170MUVH Operating Instructions ...

Page 2: ...TTER CULTIVATION Precautions for cultures P 31 Preventing contamination P 32 CORRECT OPERATION P 33 BASIC OPERATION ON LCD TOUCH PANEL P 34 BASIC PARAMETERS Numerical input to input window P 36 Setting temperature CO2 density O2 density and high limit temperature alarm P 37 Setting key lock P 39 Removing key lock P 41 Automatic N2 gas O2 gas cylinder changeover P 43 Manual N2 gas O2 gas supply lin...

Page 3: ... P 76 Setting User ID P 76 Setting auto lock P 78 Using unlock key P 81 Removing auto lock P 82 GAS AUTO CHANGER OPTION P 83 Connecting CO2 gas cylinder P 83 Automatic CO2 gas supply line changeover P 84 Manual CO2 gas supply line A B changeover P 86 STD GAS AUTO CALIBRATION KIT OPTION P 87 ROUTINE MAINTENANCE P 90 ALARMS SAFETY AND SELF DIAGNOSIS P 91 TROUBLESHOOTING P 95 DISPOSAL OF UNIT P 98 SP...

Page 4: ...ting instructions are subject to change without notice for improvement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or the page order is incorrect Contact our sales representative or agent if any point in the operating instructions is unclear or if there are any inaccuracies No part of the operating instructions may be repr...

Page 5: ... Symbol shows This symbol means caution This symbol means an action is prohibited This symbol means an instruction must be followed Be sure to keep the operating instructions in a place accessible to users of this unit Label on the unit This mark is labeled on the cover in which the electrical components of high voltage are enclosed to prevent the electric shock The cover should be removed by a qu...

Page 6: ... of accidental breakage and safe disposal options visit ec gc ca mercure mercury Dispose of or recycle in accordance with applicable laws Pour plus de renseignements sur les procédures de manutention sécuritaire les mesures à prendre en cas de bris accidentel et les options d élimination sécuritaire visitez ec gc ca mercure mercury Mettez au rebut ou recyclez conformément aux lois applicables For ...

Page 7: ...ower supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit Connect the unit to a power source as indicated on the rating label attached to the unit Use of any other voltage or frequency other than th...

Page 8: ... Disconnect the power supply plug before moving the unit Take care not to damage the power supply cord A damaged cord may cause electric shock or fire Disconnect the power supply plug when the unit is not used for long periods Keeping the connection may cause electric shock current leakage or fire due to the deterioration of insulation If the unit is to be stored unused in an unsupervised area for...

Page 9: ...ay be damaged due to the change of setting Be careful not to tip over the unit during movement to prevent damage or injury Prepare a safety check sheet copy the last page when you request any repair or maintenance for the safety of service personnel Wear rubber gloves when handling the H2O2 reagent Direct contact with the H2O2 reagent may result in inflammation of the skin H2O2 decontamination can...

Page 10: ...rface This symbol indicates an earth This symbol means ON for a power switch This symbol means OFF for a power switch ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions based on the IEC 61010 1 Indoor use Altitude up to 2000 m Temperature 5o C to 40o C Maximum relative humidity 80 for temperature up to 31o C decreasing linearly to 50 relative hum...

Page 11: ...stem set MCO 170UVS is installed 2 MCO 170MUVH or when an optional electric lock MCO 170EL is installed 23 When some inner attachments are removed 24 25 Access port cover Silicon cap Rear left When 16 is removed Sample air outlet cap 17 18 19 1 20 21 22 Handles Leveling feet Leveling foot Lower right side Removable power supply cord port ...

Page 12: ...the gas injection nozzle tube Refer to page 18 for installing 11 Humidifying pan cover This cover prevents the UV light entering the chamber Always use it Using without it may have a bad influence on the chamber temperature distribution and humidity recovery 12 Gas injection nozzle Refer to page 18 for installing 13 Humidifying pan Fill the humidifying pan with sterile distilled water and set the ...

Page 13: ...een will appear when the power switch is turned ON Note It takes approximately 20 seconds until Top screen is displayed During warming up Status Gas sensor initializing is displayed in the Message display field 18 and is displayed in the present CO2 density display field 5 and the present O2 density display field 10 2 4 3 1 1 Present temperature display field The current chamber temperature is dis...

Page 14: ...cylinder is empty Changeover is also workable by pressing this key 10 Present O2 density display field The current chamber O2 density is displayed Nothing is displayed when O2 density is set 20 11 Set O2 density value display field The set value of the chamber O2 density is displayed Default setting 20 12 N2 O2 gas injection indicator This lamp lights when N2 gas O2 gas is being injected 13 N2 O2 ...

Page 15: ...or Locked is displayed 2 18 Message display field Alarms errors or messages are displayed when fault occurs Refer to page 91 94 Note When there are a number of alarms errors the display shows the message For example if 2 alarms errors occur in total the display shows 1 2 19 Message select key When there are a number of alarm errors the message on the screen is changeable 20 Menu key Press this key...

Page 16: ...inal The alarm of this product can be informed at a remote location from this product by connecting the external alarm unit to the remote alarm terminals For the type and behavior of remote alarm output refer to page 91 94 The terminal of the remote alarm is installed at the rear upper right of the unit See the figure on the point The alarm is outputted from this terminal Contact capacity is DC 30...

Page 17: ...s boilers ovens or autoclaves Heat will adversely affect the performance of the incubator Ambient temperature at least 5 C lower than set temperature The control temperature of the incubator is at least 5 C higher than the ambient temperature For example if the chamber is controlled at 37 C the ambient temperature must be 32 C or less Do not allow the ambient temperature to become too high Strong ...

Page 18: ...r In addition do not install it near water or steam pipes Moisture can cause the insulation to deteriorate which may result in current leakage or electric shock No inflammable or corrosive gas Never install the incubator in a location where it will be exposed to inflammable or corrosive gas Doing so may result in explosion or fire In addition insulation may deteriorate due to corrosion of protecti...

Page 19: ...the cable tie banding the power supply cord Prolonged banding may cause the corrosion of the cord coating WARNING Do not leave the plastic wrapping bags within reach of children If the bag is placed over a child s head it can block the mouth and nose and cause suffocation 2 Connecting a gas injection nozzle In the case of control of O2 density in the chamber connect the gas injection nozzle to the...

Page 20: ...ottom side of each gastight split door Fig 3 5 Adjust the leveling feet Adjust the leveling feet by turning them counterclockwise to level the incubator Fig 4 Note Incubating on a leaning tray may have a bad influence on the cultivation Humidifying pan Humidifying pan cover Fig 2 Gastight split door At the same level as bottom side of each gastight split doors Inner door Fig 3 Fig 4 Leveling feet ...

Page 21: ... installed by qualified personnel If the incubator is not grounded it may result in electric shock Never connect the ground wire to a gas pipe water pipe lightning rod or telephone ground wire Doing so may cause electric shock Installing a ground fault circuit breaker If using the incubator in the location with moisture or humidity cannot be avoided then it is recommended that a ground fault circu...

Page 22: ... incubator other than this product use the stacking plate MCO 170SB Refer to table 13 on page 100 When the incubator is not in use Empty the water from the humidifying pan and remove moisture from the chamber Make sure that the chamber is completely dry before closing the doors Failure to do so may result in damage Before moving the incubator Before moving the incubator empty the water from the hu...

Page 23: ...sing 2 tube bands that is provided connect the gas tube tightly to prevent it from coming off Make sure that the gas tube is not folded If the CO2 gas is supplied to multiple CO2 incubators from a single gas cylinder a CO2 solid will be formed in the gas regulator The gas regulator safety valve will operate and there may an explosive sound 3 After connecting the gas tube make sure that no gas is l...

Page 24: ...lator rated at 25 MPa G 250 kgf cm2 G 3600 psi G for the primary side and 0 2 MPa G 2 kgf cm2 G 30 psi G for the secondary side 2 Using a gas tube that is provided connect the connecting port A for N2 O2 gas pipe and the gas regulator of the N2 gas cylinder A or O2 gas cylinder A 3 Using a gas tube that is provided connect the connecting port B for N2 O2 gas pipe and the gas regulator of the N2 ga...

Page 25: ... pipe is changed A to B automatically Note The gas lines connected to the incubator will degrade over time If any deterioration or abnormalities are found during inspection replace the lines immediately WARNING O2 gas increases the susceptibility of substances to burn Take care of the handling of flame in a room where the incubator is installed ...

Page 26: ...ting Be careful to keep the diluted detergent or water out of the temperature sensor the CO2 gas injection port the gas injection port the inner sample air access port the fan motor shaft bearing and the inner sample air outlet Fig 2 Also do not wash the temperature sensor and the UV lamp using detergent Doing so may cause failure Fig 2 3 Soak a gauze or unwoven cloth in distilled water and wring ...

Page 27: ... 170MUVH 170MUV or when an optional UV system set MCO 170UVS is installed 1 Turn OFF the power to the incubator 2 Open the outer and inner doors and pull out all the trays Fig 1 Note Open the inner door after closing all the gastight split doors 3 Pull out the humidifying pan cover Fig 2 4 Pull out the N2 O2 gas injection nozzle and N2 O2 gas injection nozzle tube Fig 3 Fig 3 Humidifying pan cover...

Page 28: ...l out the humidifying pan Fig 4 6 Pull out the fan cover Fig 5 7 Lift the duct and remove it from the pins on the rear side Fig 6 8 Pull out the fan Fig 7 Fig 5 Fig 6 Humidifying pan Duct Fan cover Fig 7 Fig 4 ...

Page 29: ...28 BEFORE COMMENCING OPERATION 9 Remove the silicon caps of each access port from interior Fig 8 and exterior Fig 9 Silicon cap Silicon cap Fig 9 Fig 8 ...

Page 30: ...over position the long hole of duct with the projection of fan cover and insert directly Fig 3 Same applies for the humidifying pan cover The fan cover may lean by strongly pushing the head of it in the back Make sure that there is no space in the lower fan cover after installing because the leaning fan cover may have a bad influence on the camber temperature distribution CAUTION If the fan cover ...

Page 31: ...inal direction as its shorter side is placed in the back Refill the humidifying pan with water early when the volume of water is decreased Mixing any reagent in the humidifying water may have a bad influence on the cultivation Especially when using the UV lamp do not use any reagent Because the UV light may deteriorate the reagent mixed with the humidifying water 5 Set the humidifying pan with the...

Page 32: ... culture medium incomplete closing or damage to the gasket Before opening the inner door or some gastight split doors check through the glass to confirm that the UV lamp is OFF if the MCO 170MUVH 170MUV or the optional MCO 170UVS is installed Be careful when closing the outer door Use the handle when closing the outer door Holding the door in other places may cause injury by getting fingers caught...

Page 33: ...aking them in and out Keep the chamber clean Wipe off any fingerprints If water spills from the humidifying pan or if the doors are left open for a long time condensation may form on the inside of the gastight split doors If that occurs wipe off the condensation with dry sterile gauze In particular clean and disinfect the chamber if the culture medium is spilled For details refer to ROUTINE MAINTE...

Page 34: ...ncubator 7 MCO 170MUVH 170MUV or when an optional UV system set MCO 170UVS is installed Set the frequency of a power supply on the LCD touch panel Refer to page 66 67 WARNING Always use the removable power supply cord that is provided Other power supply cord may cause electric shock or fire The provided removable power supply cord is only for this product Never use it for any other products When t...

Page 35: ...arm Display Alarm log can be exported 57 58 Alarm Export Export Alarm log 59 61 Lock Lock screen Auto Lock Auto Lock screen 1 Setting Auto lock ON OFF User ID ON OFF 78 82 KeyLock Key Lock screen Setting Key lock ON OFF password 39 42 Auto Lock User Auto Lock User screen Setting User ID password 76 78 Tools 1 Tools 1 screen Temp Gas Calibration Don t push See Next page Alarm Setting 1 Setting Temp...

Page 36: ...line select key Operate Manual changeover between CO2 gas supply line A and B Page 86 N2 O2 gas supply line select key Operate Manual changeover between N2 O2 gas supply line A and B Page 45 Message select key Operate Changing some messages H2O2 key 5 Operate H2O2 decontamination Page 70 75 Unlock key 6 Operate Unlock the outer door Page 80 Buzzer key Operate Silencing the buzzer Alarm is not canc...

Page 37: ...ue and press OK key Key description Numerical key 0 9 Input numerical values Up Down key Increases or decreases the numerical value displayed in the numerical input box Clear key Deletes the numerical value displayed on the numerical input box Cancel key Stops inputting on the numerical input box and closes the input window Note While the input window is open it is not possible to operate Top key ...

Page 38: ...CO2 density the O2 density and the temperature of the high limit temperature alarm for normal operation according to the following procedure The incubator automatically starts operation using these settings after power on 1 Press Menu key to lead the Menu screen 2 Press Set key to lead the Stand by Setting screen ...

Page 39: ...etting 53 0 o C Refer to page 91 94 for detail of each alarm Note When changing the set temperature from less than 45 0 o C to 45 0 o C or higher the incubator readjusts the CO2 sensor and O2 sensor During readjusting Status Gas sensor initializing is displayed in the Message display field and is displayed in the Present CO2 density display field and the Present O2 density display field After 15 m...

Page 40: ...39 Setting key lock 1 Press Menu key to lead the Menu screen 2 Press Lock key to lead the Lock screen 3 Press KeyLock key to lead the Key Lock screen ...

Page 41: ...ight Key Lock turns to ON Password The number Max 6 digit inputted here are registered the release password of Key Lock Confirm Password To prevent erroneous input input the same password as Password input box When inputting a different password Notice dialog box is displayed Press OK key and input the correct password Note To prevent abuse of the release password of Keylock manage properly by lim...

Page 42: ...displayed and input of the release password of Key Lock is required When the inputted password is incorrect Notice dialog box is displayed Press OK key and then input the correct password Removing key lock 1 By pressing Menu key the Password input window is displayed 2 On Password input box input the set release password of Key Lock and press OK key to lead the Menu screen ...

Page 43: ... Lock screen 5 On the Key Lock screen by holding Key Lock slide key and sliding to the left change to OFF Press Apply key to turn the key lock OFF The display returns to the Lock screen Note The release password of key Lock is deleted 6 On the Lock screen press Top key to return to the Top screen ...

Page 44: ...to changer MCO 21GC which changes the empty CO2 gas cylinder into the other one automatically is available as an optional accessory refer to page 83 86 Refer to page 23 24 for connecting N2 gas cylinder O2 gas cylinder N2 O2 gas supply line indicator A or B being used is lighted in the Top screen N2 O2 gas supply line indicator A and B N2 O2 gas supply line select key N2 O2 gas supply line being u...

Page 45: ...mpty it changes into N2 gas cylinder A Table 2 N2 O2 gas supply line automatic changeover e g When N2 gas cylinder A is empty it changes over N2 gas cylinder B Situation N2 gas N2 O2 gas supply line indicator Message display field Supply line Cylinder A Cylinder B A B 1 N2 gas is supplying from valve A A Remaining Remaining Light on Light off 2 O2 density in the chamber is on the increase even if ...

Page 46: ... line A is changed to B by N2 gas O2 gas automatic changer function the N2 O2 gas supply line B is changed to A manually without pressing the Buzzer key Table 3 Situation N2 gas N2 O2 gas supply line indicator Message display field Supply line Cylinder A Cylinder B A B 1 N2 O2 gas supply line A is changed to B automatically B Empty Remaining Reverse video and blink Light on Err02 N2 O2 Gas Empty a...

Page 47: ...lead the Alarm Setting 1 screen it is possible to set automatic set temperature alarm automatic set CO2 density alarm automatic set O2 density alarm and alarm delay of each alarm go to procedure 4 on page 47 Press Alarm Setting 2 key to lead the Alarm Setting 2 screen it is possible to set door alarm delay ring back and remote alarm go to procedure 5 on page 48 ...

Page 48: ...Settable range 0 5 5 0 factory setting 1 0 Automatic set CO2 density alarm delay CO2 Alarm Delay The function is that when the incubator is in the alarm state of automatic set CO2 density the alarm buzzer will sound after the automatic set CO2 density alarm delay set time passed Settable range 0 minute 15 minutes factory setting 15 minutes Note When the incubator is recovered from the alarm state ...

Page 49: ...0 minutes Note At Err01 CO2 gas cylinder empty Err02 N2 O2 gas cylinder empty Err11 12 CO2 sensor error Err18 UV lamp failure and Door alarm the alarm is not re activated because the alarm itself is deactivated by pressing Buzzer key refer to page 91 and 93 Remote Alarm The function is that the remote alarm is continued even though the buzzer is stopped by pressing Buzzer key By holding and slidin...

Page 50: ... log data chamber temperature CO2 density O2 density and open close state of outer door Use the following procedure to set the log interval interval of acquiring the operation log 1 Press Menu key to lead the Menu screen 2 Press Log key to lead the Log screen 3 Press Setting key to lead the Setting screen ...

Page 51: ...between log interval and the estimated amount of data that can be saved Log interval 2 minutes Approx 46 days Log interval 6 minutes Approx 135 days Log interval 30 minutes Approx 664 days When saving data more than the above and the data is overwritten and the old data is delated 5 Press Top key to return to the Top screen Displaying operation log Operation log saved in the incubator can be displ...

Page 52: ...reen by pressing Show key after pressing the item you want to display graphically the graph of each operation log is displayed Actual Temp Chamber temperature log graph Go to procedure 6 Actual CO2 Level CO2 density log graph Go to procedure 7 Actual O2 Level O2 density log graph Go to procedure 8 Door Opening Open close state of outer door log graph Go to procedure 9 ...

Page 53: ...o the Chart screen Press Top key to return to the Top screen 9 Door Opening log graph is displayed Press Back key to return to the Chart screen Press Top key to return to the Top screen Note The error of about 1 minute may be observed during a month Refer to page 62 for the procedure of setting time To Actual CO2 Level log To previous day To next day To Door Opening log To previous day To next day...

Page 54: ...the USB memory into the USB port Note It is not possible to use a USB memory with security functions that requires entering password 11 Press Export key 12 When the export is complete Information dialog box is displayed Press OK key Refer to page 55 56 for the details about abnormal export or exported file name 13 Press Top key to return to the Top screen ...

Page 55: ...t to the USB memory inserted into the USB port 1 Insert the USB memory into the USB port Note It is not possible to use a USB memory with security functions that requires entering password 2 Press Menu key to lead the Menu screen 3 Press Log key to lead the Log screen 4 Press Data Export key to lead the Export screen ...

Page 56: ... log data you want to export To export all types of operation log data press All Ch key To export only operation log data you want to export select operation log data you want to export and then press Selected Ch key Actual Temp Chamber temperature log data Actual CO2 Level CO2 density log data Actual O2 Level O2 density log data Door Opening Open close state of outer door log data When both of th...

Page 57: ...er is created in the USB memory and the exported file is saved in it in CSV format The exported file name is in date 8 digits type of data format e g When exporting all types of data using All from Oct 1st 2015 to Jan 1st 2016 20151001 20160101_AllCh csv 20151001 20160101_Door csv e g When exporting Actual Temp using 1 Day Jan 1st 2016 20160101_Temp csv On the beginning of the exported file MCO 17...

Page 58: ...s Alarm log saved in the incubator can be displayed graphically on the LCD touch panel Note When saving alarm logs more than 257 the oldest alarm log is deleted and then overwritten 1 Press Menu key to lead the Menu screen 2 Press Log key to lead the Log screen 3 Press Alarm key to lead the Alarm screen ...

Page 59: ... range 1 day 45 days Note The error of about 1 minute may be observed during 1 month Refer to page 62 for the procedure of setting time Press Back key to return to the Log screen Press Top key to return to the Top screen On the Alarm screen of procedure 4 or 5 alarm log data can be exported in CSV format to the USB memory inserted into the USB port 6 Insert the USB memory into the USB port Note It...

Page 60: ... the USB port by CSV format 1 Insert a USB memory in the USB port Note It is not possible to use a USB memory with security functions that requires entering password 2 Press Menu key to lead the Menu screen 3 Press Log key to lead the Log screen 4 Press Alarm Export key to lead Alarm Export screen ...

Page 61: ... button and input days Settable range 1 day 45 days Note The error of about 1 minute may be observed during 1 month Refer to page 62 for the procedure of setting time 6 Press Export key Note When USB memory is not inserted in the USB port Notice dialog box is displayed Press OK key and insert an USB memory into the USB port When alarm log data doesn t exist in the specified days Notice dialog box ...

Page 62: ...t deleted 8 Remove a USB memory from the USB port Note A log folder is created in a USB memory and an exported data file is saved in the log folder by CSV format Exported file name The first date during exported period 8 digits the last date 8 digits AlarmLog Example When exporting alarm log data for 7 days on January 7 2016 20160101 20160107_AlarmLog csv 9 Press Top key to return to the Top scree...

Page 63: ... key to lead the Date Time screen 4 On the Date Time screen input the present date and time Press Apply key to save the input value The display returns to the Tools 2 screen Note 24 hour clock It is recommended to set the time periodically since the error of about 1 minute may be observed during a month 5 Press Top key to return to the Top screen ...

Page 64: ...s Tools 2 key to lead the Tools 2 screen 3 Press Brightness Sleep key to lead the Brightness Sleep screen 4 On the Brightness Sleep screen each setting of brightness and sleep is available Press Apply key to save the input value and setup The display returns to the Tools 2 screen ...

Page 65: ... right the Sleep function is turned to ON Input the set value of time to change the Sleep state Settable range 1 minute 5 minutes factory setting 2 minutes Note It is not possible to operate any key in the Sleep state By touching the LCD touch panel the Sleep state is released and the LCD touch panel returns to the usual state Under this condition key operations are available Brightness Sleep Brig...

Page 66: ...he condensation in the chamber and chamber temperature distribution because the UV lamp generates heat for a long time It may also shorten the service life of the UV lamp The preset period can be changed when necessary as shown in the page 66 67 To check whether the UV lamp is lit open the outer door and then press the door switch with the inner door close Visible blue light can be confirmed from ...

Page 67: ...r information on replacing the UV lamp If the UV lamp burns out Err18 UV Lamp Abnormal is displayed in the message display field it is not possible to perform H2O2 decontamination Replace the UV lamp and the glow starter Setting UV lamp ON period Use the following procedure to change the setting of the UV lamp ON period 1 Press Menu key to lead the Menu screen 2 Press Tools 1 key to lead the Tools...

Page 68: ... lamp has turned on is displayed as the percentage to 5 000 hours which are recommendation time to replace It is impossible to set UV Timer Ext The more total time which UV lamp has turned on increases the more UV ray output declines In order to cover a decline of the UV ray output the lighting time of UV lamp is automatically extended with an increase of total lighting time of UV lamp The set val...

Page 69: ...n cover the duct the fan the humidifying pan the humidifying pan cover the gas injection nozzle and the gas injection nozzle tube Clean all the attachments in an autoclave or with alcohol for disinfection 2 Clean and wipe off inside the chamber with alcohol for disinfection 3 Set the CO2 density to 0 and the O2 density to 20 Refer to page 37 38 4 Press Menu key to lead the Menu screen 5 Press Tool...

Page 70: ...on display when UV lamp is lighting Note The UV 24 hour mode may cause the automatic set temperature alarm because of a rising chamber temperature After procedure 8 by opening the outer door when UV lamp is lighting UV lamp is turned OFF and UV 24 hour mode is canceled by opening the outer door Redo from procedure 4 to start the UV 24 hour mode again 9 Press Top key to return to the Top screen 10 ...

Page 71: ... result in damage to the H2O2 vapor generator WARNING When performing H2O2 decontamination make sure that the gastight split doors the inner door and the outer door are securely closed During H2O2 decontamination plug the access hole with the silicon caps that are provided Failure to do so may be harmful to health due to leakage of H2O2 gas CAUTION H2O2 decontamination can be performed only for th...

Page 72: ...e H2O2 generator in the 2 holes on the lower left side of the duct Fig 1 Note Make sure that the H2O2 generator is securely attached Uncertain attachment may result in insufficient decontamination 7 Remove a connector cap from connector on bottom right of the far side of the chamber Connect the H2O2 generator and connector by a cable that is provided Fig 2 Note Be sure to keep the connector cap 8 ...

Page 73: ... gastight split doors the inner door and the outer door 11 Press H2O2 key for 3 seconds to open Setting Position window Note When the model does not have H2O2 decontamination function H2O2 key is not displayed on the Top screen When key lock is ON Password input window is opened and input of the release password of Key Lock is required Refer to page 41 12 Finally make sure that the inner attachmen...

Page 74: ...uzzer sounds when H2O2 decontamination is completed About 100 minutes later CAUTION The electric lock will remain locked if power supply is cut off during H2O2 decontamination After the power supply is recovers the H2O2 gas resolution process will start execution and finish automatically Execute the decontamination again because the decontamination is not completed WARNING Do not use the unlock ke...

Page 75: ...r and inner doors Disconnect connector of the chamber cable and H2O2 generator and remove H2O2 generator and cable from the chamber Note When doing the above work put on protection glasses and rubber gloves 16 On the H2O2 Decontamination Step8 screen press OK key to return to the Top screen Note When H2O2 generator is connected with connector of the chamber by cable OK key is not workable 17 Dilut...

Page 76: ...to prevent impurities from becoming mixed in the reagent Check the container to make sure that there is no damage corrosion or cracking Store the container with the inlet facing upwards and make sure that the container will not tip or be knocked over Precautions for disposal Dispose according to the rule in your country Using unlock key Unlocking when power is interrupted MCO 170MUVH or when an op...

Page 77: ...he MCO 170MUV 170M The modes of unlocking the outer door are as follows Quick mode Press the Unlock key User ID mode Input the User ID and release password of Auto Lock after pressing the Unlock key Setting User ID Before turning the User ID mode to ON use the following procedure to register an User ID and a release password of Auto Lock 1 Press Menu key to lead the Menu screen 2 Press Lock key to...

Page 78: ...word as Password input box When inputting a different password Notice dialog box is displayed Press OK key and input the correct password Note A release password of Auto Lock is for unlocking the outer door It is different from the release password of Key Lock refer to page39 42 It is possible to input up to 8 digit alphanumeric characters as a User ID It is possible to input up to 6 digit numbers...

Page 79: ... registered User ID into User ID input box and input its registered password into Password input box Press Delete key to delete the registered User ID and its password Note When deleting all registered User IDs the User ID mode is turned to OFF refer to page 79 5 On the Menu screen press Back key to return to the Top screen Setting auto lock 1 Press Menu key to lead the Menu screen 2 Press Lock ke...

Page 80: ... of unlocking the outer door between the quick mode or the User ID mode By holding the User ID slide key and sliding it right the User ID mode is turned to ON Factory setting OFF quick mode Note When no User ID is registered notice dialog box is displayed Press OK key and then register a User ID and its password in reference to page 76 78 In the User ID mode User ID which is inputted to unlock the...

Page 81: ...r ID and its release password of Auto Lock Note The User ID which is inputted at this time is saved as the open close state of outer door log data refer to page 54 56 When the inputted User ID or its password is incorrect Notice dialog box is displayed Press OK key and then input the correct User ID or its password Note When the unlocked outer door is closed and the setting time passes the unlocke...

Page 82: ...e power is interrupted use the unlock key that is provided To re lock the outer door turn the unlock key to the lock direction while the outer door is open Close the outer door after the out door is locked manually Note The outer door cannot be locked by using the unlock key while the outer door is closed Lock the outer door while it is open Attempting to turn the unlock key while the outer door i...

Page 83: ...ad the Lock screen 3 Press Auto Lock key to lead the Auto Lock screen 4 On the Auto Lock screen by holding the Auto lock slide key and sliding it left the Auto lock is turned to OFF Press Apply key to change Auto lock OFF and the display returns to the Lock screen 5 Press Top key to return to the Top screen ...

Page 84: ...as regulator of the CO2 gas cylinder A 3 Using a gas tube that is provided connect the connecting port B for CO2 gas pipe and the gas regulator of the CO2 gas cylinder B Note If the CO2 gas is supplied to multiple CO2 incubators from a single gas cylinder a CO2 solid will be formed in the gas regulator The gas regulator safety valve will operate and it may make an explosive sound 4 After connectin...

Page 85: ...to changer MCO 21GC is installed CO2 gas supply line indicator A B and CO2 gas supply line select key are displayed in the Top screen CO2 gas supply line indicator A or B being used is lighted CO2 gas supply line indicator A and B CO2 gas supply line select key CO2 gas supply line being used ...

Page 86: ...gas cylinder A Table 4 CO2 gas supply line automatic changeover e g When CO2 gas cylinder A is empty it changes over CO2 gas cylinder B Situation CO2 gas CO2 gas supply line indicator Message display field Supply line Cylinder A Cylinder B A B 1 CO2 gas is supplying from valve A A Remaining Remaining Light on Light off 2 CO2 density in the chamber is not increased even if CO2 gas valve opens A Emp...

Page 87: ... gas supply line A is changed to B by CO2 gas automatic changer function the CO2 gas supply line B is changed to A manually without pressing the Buzzer key Table 5 Situation CO2 gas CO2 gas supply line indicator Message display field Supply line Cylinder A Cylinder B A B 1 CO2 gas supply line A is changed to B automatically B Empty Remaining Reverse video and blink Light on Err01 CO2 Gas Empty and...

Page 88: ... incubator Since a Standard gas cylinder is used as a standard of exact density during CO2 O2 density calibration prepare a standard gas cylinder that is same as the set CO2 O2 density Note There is not the problem to remain connected standard gas cylinder after finished CO2 O2 gas density calibration 2 Press Menu key to lead the Menu screen 3 Press Tools 1 key to lead the Tools 1 screen 4 Press S...

Page 89: ...ting 5 0 ON O2 standard gas density O2 STD Gas Density O2 density of the connected standard gas cylinder By holding O2 STD Gas Density slide key and sliding it to the left O2 density calibration turns to OFF Turn OFF when O2 density calibration is not required Settable range 0 5 18 0 and 22 0 81 0 Factory setting 5 0 ON e g Standard gas density Calibrate both of CO2 and O2 density CO2 5 O2 5 Calib...

Page 90: ...ration starts Calibration go to Step5 Procedure 10 automatically 10 After completion of gas density calibration display leads the STD Gas Calibration Step5 screen O2 CO2 incubator returns to the normal operation 11 On the STD Gas Calibration Step5 screen press OK key to return to the Tools 1 screen On the Tools 1 screen press Top key to return to the Top screen ...

Page 91: ...ast once a month 1 Remove all the inner attachments by the procedures shown on page 26 28 2 Clean the chamber and all the inner attachments by the procedures shown on page 25 3 Install all the inner attachments by the procedures shown on page 29 When there is excessive dirt contact our sales representative or agent ...

Page 92: ...for approx 90 s Return to the Top screen Door alarm The outer door is open Door Open is displayed alternately in normal characters and reverse video in the outer door opening closing display Intermittent tone after set alarm delay time 1 min to 30 min has elapsed The CO2 valve and N2 O2 valve are closed The heater turns OFF after 1 min Door lock error MCO 170MUVH or when an optional MCO 170EL is i...

Page 93: ...Door Heater Abnormal is displayed in the message display field Sensor box heater error 1 a High limit temperature alarm is activated b The sensor box heater burnout occurs or the sensor box heater SSR is short circuited Err16 CO2 Box Heater Abnormal is displayed in the message display field Heater SSR burnout 1 a High limit temperature alarm is activated b Main bottom door or sensor box heater SSR...

Page 94: ...When the Ring Back set time passes ON Non interlock with Buzzer key ON ON ON Under continuation ON Continue ON Under continuation OFF OFF Interlock with Buzzer key ON OFF Note Close the outer door when Err38 is activated Table 9 In the case of Err01 CO2 gas cylinder empty Err02 N2 O2 gas cylinder empty Err11 12 CO2 sensor error Err18 UV lamp failure or door alarm 2 Remote Alarm setting Ring Back s...

Page 95: ...nuous tone when the outer door is open ON when the outer door is open H2O2 mist generation is cancelled Resolve time is extended Failure during H2O2 gas resolve by UV The UV lamp failed during H2O2 gas resolve by UV Err36 UV Lamp Failure is displayed in the message display field Intermittent tone ON Resolve time is extended Power was interrupted during H2O2 gas resolve by UV After power is restore...

Page 96: ... G or lower it is a sign that there is little gas remaining Replace the cylinder soon Is the incubator operating beside the appliance that generates the electromagnetic wave The chamber temperature does not match the set value Is the ambient temperature less than 5 C different from the set value for the chamber temperature Is the outer door closed with the inner door or the gastight split doors le...

Page 97: ...hat the HEPA filter may be clogged Contact our sales representative or agent Is there little gas remaining in the gas cylinder Is the secondary pressure for the gas regulator of CO2 gas cylinder at the specified value of 0 1 MPa G 1 kgf cm 2 G 14 5 psi G Is the secondary pressure for the gas regulator of N2 gas O2 gas cylinder at the specified value of 0 1 MPa G 1 kgf cm 2 G 14 5 psi G Is the gas ...

Page 98: ...fan attached properly Confirm if the fan is pushed all the way to the motor shaft Fig 2 Is the H2O2 generator securely installed Set the 2 pins of the H2O2 generator in the 2 holes at the left bottom of the duct Fig 3 Is it the end of life of the H2O2 generator If the total operating time exceeds 5 000 hours replace the H2O2 generator Fig 1 Fig 2 Fig 3 Note If the problem still has not been solved...

Page 99: ...r must be dismantled and disposed of by qualified personnel only If the O2 CO2 incubator is left where outsiders enter it may result unexpected accident for example children to become locked inside Before disposing the O2 CO2 incubator with biohazardous danger decontaminate the O2 CO2 incubator to the extent possible by the user ...

Page 100: ...idifying pan Temperature controller PID control system Temperature display Digital display CO2 controller PID control system CO2 density display Digital display O2 controller PID control system O2 density display Digital display Air circulation Fan assisted Air filter 0 3 m Efficiency 99 97 or higher UV lamp 4 W x 1 ozone free emission Alarms Automatic set temperature alarm Automatic set CO2 densi...

Page 101: ...the updated catalog when ordering an optional component Designs and specifications are subject to change without notice Table 12 Optional accessories for each function MCO 170M MCO 170MUV MCO 170MUVH To disinfect by UV UV system set MCO 170UVS standard equipment To decontaminate by H2O2 UV system set MCO 170UVS H2O2 generator MCO HP H2O2 decon board MCO 170HB Electric lock MCO 170EL H2O2 generator...

Page 102: ...ad O2 control range 1 to 18 and 22 to 80 O2 variation 0 2 ambient temperature 23 o C setting 37 o C CO2 5 O2 5 no load Chamber humidity 95 R H 5 R H Applicable environment condition Temperature 5 o C to 35 o C Humidity 80 R H max The designed performance may not be obtained If the ambient temperature is equal or less than 15 o C Noise level 25 dB A scale Power consumption Max 370 W Max 375 W Heat ...

Page 103: ... the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see below Procedure to be adhered to in order to reduce safety risk indicated in b below Date Signature Address Division Telephone Product name O2 CO2 incubator Model No MCO Serial number Date of Installatio...

Page 104: ...Printed in Japan LDCL010700 0 S0116 0 1 1 1 Sakata Oizumi Machi Ora Gun Gunma 370 0596 Japan Panasonic Healthcare Co Ltd 2016 ...

Reviews: