87
STD GAS AUTO CALIBRATION KIT (OPTION)
When a STD gas auto calibration kit MCO-SG is installed, by connecting standard gas cylinder for
calibration, it is possible to calibrate CO
2
/O
2
density manually.
1.
Connect a standard gas cylinder to connecting
port for standard gas cylinder on lower right side of
the O
2
/CO
2
incubator. Since a Standard gas
cylinder is used as a standard of exact density
during CO
2
/O
2
density calibration, prepare a
standard gas cylinder that is same as the set
CO
2
/O
2
density.
Note:
There is not the problem to remain
connected standard gas cylinder after finished
CO
2
/O
2
gas density calibration.
2.
Press Menu key to lead the Menu screen.
3.
Press Tools #1 key to lead the Tools #1 screen.
4.
Press STD Gas Setting key to lead the STD Gas
Setting screen.
Lower right side
Connecting port for standard gas cylinder