1.4 Other Specifications
Page 1-27
Setting procedures of vacuum sensor (Example: For CM301)
Make the setting on each beam. (Two indicators need to be set.)
<Reference vacuum uptake value check>
1.
Attach 1003 nozzles to each head.
∗
Use the nozzles without orifices.
2.
Select the channels (1ch, 3ch) to be set,
using
and
keys.
Each channel’s function
1ch
Head
A
Nozzle-off detection
2ch
Head
A
Pickup miss
detection
3ch
Head
B
Nozzle-off detection
4ch
Head
B
Pickup miss
detection
3.
Press
for three seconds or more.
•
When it is reset, 0 is indicated.
•
It is carried out with a vacuum OFF.
4.
Turn ON a vacuum in the output check.
5.
Read the indication on the vacuum
sensor.