DigitalFlow™ XMT868i Service Manual
9
Chapter 1. Calibration
1.7.2
Testing the Output (cont.)
IMPORTANT:
The procedure for testing all the outputs is the same. However, when testing another output, be sure to
reconnect the pulse counter at terminal block
J2
. See Figure 4 on page 12 for the correct pins.
4.
Use the arrow keys to enter a value for the minimum pulse on time (between 1
sec and 10,000
sec) for the
frequency of the totalizer pulses and press
[Enter]
.
Note:
A complete pulse consists of equal amounts of
ON
and
OFF
times. Choose a value that is compatible with
the pulse counter to be used.
5.
Use the arrow keys to enter a value for the number of measurement units represented by each pulse (between 1
and 10,000) and press
[Enter]
. That number of pulses will then be output at the specified frequency.
6.
Verify that the pulse counter reads the correct value.
1.7.2.1
Procedure Options
This completes testing of the
Slot 1
totalizer outputs. Do one of the following:
• To test another of the available totalizer outputs, return to Step 3 in
Testing the Output
above. If any of the
totalizer outputs fails the test, contact the factory for assistance.
• To calibrate/test additional inputs/outputs, proceed to the appropriate section.
• To leave the
Keypad Program
, press
[Escape]
twice.
Summary of Contents for DigitalFlow XMT868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 24: ...Chapter 2 Error Codes 4 DigitalFlow XMT868i Service Manual...
Page 34: ...Chapter 3 Diagnostics 10 DigitalFlow XMT868i Service Manual...
Page 44: ...Chapter 4 Parts Replacement 10 DigitalFlow XMT868i Service Manual...
Page 48: ...Chapter 4 Parts Replacement DigitalFlow XMT868i Service Manual...
Page 52: ...Appendix A Service Record 18 DigitalFlow XMT868i Service Manual...
Page 66: ...Appendix B Calibration and Testing with PanaView 32 DigitalFlow XMT868i Service Manual...
Page 72: ...Warranty 38 DigitalFlow XMT868i Service Manual no content intended for this page...
Page 73: ......