Appendix D. Foundation Fieldbus Communications
130
DigitalFlow™ DF868 Programming Manual (2-Channel)
D.2 Configuration Utility Setup
The following is an example setup using National Instruments Configuration Utility v3.1.
Figure 40 shows the Configuration Utility with a flowmeter on the network (Panametrics Flow-XMT).
Figure 40: Configuration Utility Setup Example
Note:
The following procedures assume that the device has been placed in the OOS (out-of-service) mode before
executing.
D.3 Selecting the Desired Measurements
To set the measurement unit for each AI:
1.
Double click on the
FLOW
Transducer Block (in the tree under
GEFlow-XMT
).
2. Select the
Others
tab and open the drop down list for the
PRIMARY_SELECTOR
and
SECONDARY_SELECTOR
(refer to
3. Choose the unit from the list (see Figure 41 on page 131).
This unit will correspond to the unit that is available in the AI block for network connection. The
PRIMARY_SELECTOR
unit
will correspond to
ANALOG_INPUT_1
and the
SECONDARY_SELECTOR
will correspond to
ANALOG_INPUT_2
.
4. After the desired measurements have been selected for the
PRIMARY
and
SECONDARY SELECTOR
, choose the unit
system (
UNIT_SELECTOR
above the
PRIMARY_SELECTOR
) that has been programmed in the flowmeter (English or SI).