![Panametrics C-RL Instruction Manual Download Page 9](http://html1.mh-extra.com/html/panametrics/c-rl/c-rl_instruction-manual_3404851009.webp)
Model C-RL Installation Guide
1
1
Introduction
The C-RL clamp-on ultrasonic flow transducers are used exclusively with the Panametrics Measurement & Control
line of ultrasonic flowmeters. These transducers are used to measure the flow of various gases through pipes having
diameters from 3/4 in. (20 mm) to 30 in. (750 mm). Such measurements are independent of the pipe material. This
manual provides details on the following topics:
• Transducer construction
• Dampening material
• Installation
• Maintenance
• Specifications
2
Transducer Construction
A transducer assembly includes a 316 stainless steel housing with a plastic wedge attached and a BNC-type
electrical connector. The transducer features both integral 3/4” NPT-M threads for mounting a junction box and
integral 5/8-24 UNEF threads for mounting a submersible connector if desired (see
Figure 1 below).
Figure 1: Transducer Assembly (Ref. dwg. 752-163)
5/8-14 UNEF Thread
Stainless Steel Housing
3/4” NT Thread
Plastic
BNC Connector
3”
4”
Wedge