Getting started with the module for OEM   

Getting started with the module for OEM 

1

 

Introduction 

The EmStat-series are potentiostats with an embedded microcontroller. They provide all the major 
potentiostatic techniques with automatic current ranging and peripheral control.  
 

1.1

 

EmStat communications 

The EmStat is interfaced to a PC or host controller by means of a powered USB bus or serial (TTL) port. 
The instrument can be used with the Windows program PSTrace.  
 

 

 
Several Software Development Kits (SDKs) for .NET platforms are available to quickly and easily develop 
your application for EmStat. The SDK

s allow you to: 

 

easily connect to instruments; 

 

run measurements; 

 

store data in PSTrace / PStouch compatible format; 

 

peak detection; 

 

equivalent circuit fitting; 

 

manually control the cell; 

 

and many other things. 
 

SDK’s are available for

 

WinForms 

 

WPF 

 

Android (Xamarin) 

 

UWP 

 
All SDK

’s 

can be downloaded from 

my.palmsens.com

 

 

The instrument can also be controlled by using the commands as described in the ‘Communications 
protocol’

 document. 

 

1.2

 

USB or serial / TTL 

 is normally controlled and power via its mini USB port. This requires the PalmSens drivers to 
be installed (automatically installed with PSTrace or the .NET SDK).  
Optionally the USB port can be re-programmed to function as a virtual COM port, see appendix A. 
 

 
 can also be controlled from a host controller by means of Tx and Rx (5V TTL). Use the 
following pins for TTL communications (see next page for pin locations): 
CON2 - pin 12:  +5V power supply (500 mA min.) 
CON2 - pin 11:  GND (or CON2 

 pin 6) 

CON2 - pin 1:   Tx 
CON2 - pin 2:   Rx 
 
 

Make sure the USB chip is not programmed as  

VCP (Virtual COM port) when using the Rx/Tx lines on CON2. 

Summary of Contents for EmStat3+

Page 1: ...Getting started with the EmStat3 module for OEM OEM Interface for electrochemical sensors Last revision January 28 2020 2020 PalmSens BV www palmsens com...

Page 2: ...r your convenience and may be superseded by updates It is your responsibility to ensure that your application meets with your specifications PALMSENS BV MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY K...

Page 3: ......

Page 4: ...at communications 3 1 2 USB or serial TTL 3 1 3 Baudrate settings 4 1 4 Connecting to PC via RS232 converter 5 2 Printed Circuit Board Layout 5 CON1 description 6 2 1 CON2 description 6 2 2 Heat trans...

Page 5: ...n equivalent circuit fitting manually control the cell and many other things SDK s are available for WinForms WPF Android Xamarin UWP All SDK s can be downloaded from my palmsens com The instrument ca...

Page 6: ...r measurements Baudrate Max number of data points second Min interval time 230400 1000 1 00 ms 57600 197 5 07 ms It is possible to change the baudrate setting of the EmStat module in the firmware requ...

Page 7: ...ircuit Board Layout EmSta3 PCB layout The board has two pin headers CON1 and CON2 CON1 is used to connect for instance the MUX multiplexer or any other peripheral CON2 is used when serial TTL communic...

Page 8: ...powered from USB 5 V power supply 100 mA min 6 DGND 7 10 V analog adviced max 50 mA 8 Sense 9 AGND connect to metal housing cable shielding 10 WE 11 RE 12 CE The digital I O lines are at 5V 2 1 CON2...

Page 9: ...a thermal connection from the bottom side of the PCB to something which can absorb heat for instance using the thermal pad WX 3009 0227 6 from 3M Suggested heatsinks Heatsink 24 pin DIP Glue On BLK 5...

Page 10: ...N4 on page 5 Front view of socket Solder side of socket 1 4 2 3 4 1 3 2 1 RE blue connector or reference electrode 2 CE black connector or counter electrode 3 S red female male connector or sense 4 WE...

Page 11: ...supply from EmStat Connect CON2 Pin 8 download active low to CON2 Pin 6 GND Connect the power supply Run program Update firmware Press Connect Load firmware file ESx where x is the EmStat type and th...

Page 12: ...ck the Scan and Parse button again to check if the settings are written correctly especially the invert DTR setting If this setting cannot be changed please contact us Your FT_PROG version might have...

Reviews: