background image

 

Getting started with the module for OEM   

Getting started with the module for OEM 

3

 

Firmware 

The firmware of EmStat can be updated 

by using the program ‘Update firmware’

 or using PSTrace for 

WIndows. The EmStat with USB interface is updated automatically.  
 

In order to follow the steps below first solder on two small wires to both pin 6 and  

pin 8 of CON2 of the EmStat module directly so you can connect the two pins briefly. 

 

 
When the EmStat is connected by means of a serial connection, it is required to toggle pins on CON2 
manually.   
Follow these steps: 

-

 

Remove the power supply from EmStat 

-

 

Connect CON2-Pin 8 (download - active low) to CON2-Pin 6 (GND).  

-

 

Connect the power supply 

-

 

Run program ‘Update firmware’ 

 

-

 

Press ‘Connect’

 

-

 

Load firmware file ESx##, where x is the EmStat type and ## the firmware version. 

-

 

Make sure the checkbox ‘PalmSens or EmStat is on, but does not work property' i

s checked 

-

 

Click button 'Update Firmware'. 

-

 

Remove the power supply briefly to force a reset 

-

 

Wait until updating has finished. 

-

 

Remove the connection between CON2-Pin 8 and CON2-Pin 6. 

-

 

Force another reset by briefly interrupting the power supply. 

 

4

 

Power requirements 

Power from USB port: 

-

 

Cell off 10 mA:   0.220 A 

-

 

Cell off 100 mA:  0.257 A 

-

 

Cell on 10 mA, with 10 mA output current: 0.277 A  

-

 

Cell on 100 mA, with 50 mA output current: 0.440 A 

 

 

CON1

1

12

CON2

12

CON4

1

4

2

3

IC17

Mini

USB

55.13 mm

41.0 mm

Summary of Contents for EmStat3+

Page 1: ...Getting started with the EmStat3 module for OEM OEM Interface for electrochemical sensors Last revision January 28 2020 2020 PalmSens BV www palmsens com...

Page 2: ...r your convenience and may be superseded by updates It is your responsibility to ensure that your application meets with your specifications PALMSENS BV MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY K...

Page 3: ......

Page 4: ...at communications 3 1 2 USB or serial TTL 3 1 3 Baudrate settings 4 1 4 Connecting to PC via RS232 converter 5 2 Printed Circuit Board Layout 5 CON1 description 6 2 1 CON2 description 6 2 2 Heat trans...

Page 5: ...n equivalent circuit fitting manually control the cell and many other things SDK s are available for WinForms WPF Android Xamarin UWP All SDK s can be downloaded from my palmsens com The instrument ca...

Page 6: ...r measurements Baudrate Max number of data points second Min interval time 230400 1000 1 00 ms 57600 197 5 07 ms It is possible to change the baudrate setting of the EmStat module in the firmware requ...

Page 7: ...ircuit Board Layout EmSta3 PCB layout The board has two pin headers CON1 and CON2 CON1 is used to connect for instance the MUX multiplexer or any other peripheral CON2 is used when serial TTL communic...

Page 8: ...powered from USB 5 V power supply 100 mA min 6 DGND 7 10 V analog adviced max 50 mA 8 Sense 9 AGND connect to metal housing cable shielding 10 WE 11 RE 12 CE The digital I O lines are at 5V 2 1 CON2...

Page 9: ...a thermal connection from the bottom side of the PCB to something which can absorb heat for instance using the thermal pad WX 3009 0227 6 from 3M Suggested heatsinks Heatsink 24 pin DIP Glue On BLK 5...

Page 10: ...N4 on page 5 Front view of socket Solder side of socket 1 4 2 3 4 1 3 2 1 RE blue connector or reference electrode 2 CE black connector or counter electrode 3 S red female male connector or sense 4 WE...

Page 11: ...supply from EmStat Connect CON2 Pin 8 download active low to CON2 Pin 6 GND Connect the power supply Run program Update firmware Press Connect Load firmware file ESx where x is the EmStat type and th...

Page 12: ...ck the Scan and Parse button again to check if the settings are written correctly especially the invert DTR setting If this setting cannot be changed please contact us Your FT_PROG version might have...