5
Operation and settings
5249-en-us_V1.2_11/20
15
5
Operation and settings
5.1
Transporting the device
NOTICE
Device damage due to working fluid
The device will sustain damage when penetrated by working fluid.
4
Keep the device horizontal and use it in this position
Observe the following when transporting the device:
1.
Protect the device against severe impact during transportation.
2.
The working fluid must be removed in the event of longer transport
routes.
3.
The working fluid can remain in the device for short distances when the
person carrying the device is walking.
Always hold the device horizontally when transporting it.
5.2
Setting up the device
1.
Set up the device on an even surface.
2.
Ensure that the ventilation slots on the rear of the housing are not
obstructed at any time.
5.3
Charging the device
The device has to be charged before using it for the first time or when the
battery is dead.
The display shows the charge level of the device.
4
Use only the power supply unit delivered with and belonging to the
device.
The power supply unit heats up slightly while charging.
Charge time
It takes 2.5 h for the device to charge completely.
5.4
Filling glass bottle
Note
Be aware of the minimum and maximum filling level. When the fluid level
falls below the minimum, the device will no longer function properly. When
the fluid level exceeds the maximum, it will escape from the device.
• 40 ml = minimum level
• 80 ml = maximum level
Proceed as follows to fill the glass bottle with working fluid:
1.
Unscrew the glass bottle and remove it from the device
2.
Pour 80 ml of working fluid into the glass bottle
3.
Screw the glass bottle into the device