3
Product description
5249-en-us_V1.2_11/20
11
3.2
Device overview
1
3
2
4
5
Front of device
1 Aerosol outlet
4 Display
2 Setpoint dial
5 Carry handle
3 Pump output dial
1
4
3
5
2
Rear of device
1 USB interface
4 Glass bottle for working fluid
2 Power switch
5 Laskin nozzle
3 Power supply unit connection
3.3
Functional description
The
PAG 1000
generates an aerosol with a particle concentration up to
4.7 x 10
7
particles/cm
3
. An integrated pump generates the compressed air. A
vacuum conveys the fluid to be nebulized through a Laskin nozzle. The size
and concentration of the particles generated are a factor of the fluid used as
well as of the pump settings. The volume flow rate can be set between 0.9
and 4.6 l/min.
The USB interface is intended to be used only by the manufacturer, e.g.
when performing repairs or maintenance.