-- Range of alarm temperature
When into the “F-2” mode, the range of settable alarm tempera-
ture is from 37.6℃(99.7℉) to 43.0℃(109.4℉).
-- Range of temperature difference
When into the “F-3” mode, the range of settable temperature
difference is from 5℃(9℉) to -5℃(-9℉).
Note: F-3 mode is only for professional doctor and nurse use!
--Taking Temperature in Forehead Temperature Mode (Adjusted
Mode)
1. Setup the measurement mode to “Body mode” (the setup step
see the above “2) Function Setup mode”;
2. Make sure the distance between thermometer and forehead is
3-5cm, and press the button of [Power/Scan] directly to measure
the forehead temperature .
3. When you hear one beep sound, the LCD will show the result.
NOTE:
1. When measuring, if the environment exceed the 10-40℃ range
will be showing Err;
2. If the measured temperature result is < 32.0℃, the LCD will be
showing Lo;
3. If the measured temperature result is >43.0℃, the LCD will be
showing Hi;
4. If the measured temperature result is ≥38.0℃, the buzzer will be
send out “Beep—Beep—Beep”, a long one and two short ones.
5. We recommend measurement for 3 times to reference, and keep
the interval between the two measurements by 5 minutes at least.
4)Temperature measuring
TEMPERATURE TAKING HINTS
- Before take the temperature, please fix the forehead
hair to prevent measured deviation.
- The sweat on head or cosmetic can affect the accuracy
of measurement, please maintain the cleanness of the
forehead when measuring.
- It is normal that there may be temperature difference
depending on various skin types and color, since different
skin type will reflect different voltage of infrared ray.
- Don’t use the thermometer outdoors.
- 11 -