background image

--Taking Temperature in Surface Temperature Mode (Direct Mode)

1. Setup the measurement mode to “Surface mode” (the setup

step see the above “2) Function Setup mode”;

2. Aim the probe tip at the object which will be measured.

3. Press [Power/Scan] button, then the measurement completed.

It will show the measurement data after a “bi” sound.

4. If the measured temperature is higher than 100℃ , display “Hi”.

If the measured temperature is below 0℃,display “Lo”.

5)Automatic shutdown

If you do nothing after the temperature measurement completed 

and 15 seconds later, the thermometer will automatically 

shutdown. 

6)

 Memory function

In the power-on state, press the Up/Down button into memory 

query mode, the LCD will be showing the recorded values or “ ----“ 

symbol if not record accompanied with memory M symbol. The 

KRT01

 can restore up to 99 sets data. the query interface can 

display memory serial number, body mode and recorded tempera-

ture values.

Remark: The temperature result of surface mode can not be saved. 

And the abnormal body temperature result can not be saved too

(such as Lo, Err or Hi).

7)

 Backlight function

The backlight will be according to the different measuring 

temperature to lighting different color. The detailed as below 

table:

   TEMPERATURE TAKING HINTS

- For heat-insulated object, please don’t measure the

surface.

- For hot liquid, please do not directly measure the

surface, because the hot steam fog can condense in the

sensor lens and cause the measurement deviation.

Measuring temperature 

Backlight 

Lighting time 

< 37.5

 

Green backlight 

7s 

37.5 to 37.9

 

Orange backlight 

7s 

38.0

 

Red backlight 

7s 

Remarks: 

This specification of this table only for the factory default state. If the user into the setup 

mode to setting the orange or red backing temperature, then will compliance with the setting result. 

- 12 -

Summary of Contents for KRT01

Page 1: ...Please read this manual before use Infrared thermometer KRT01 User s Manual ...

Page 2: ...ve a very close relationship When a human forehead s temperature in 36 37 it emits wavelength 9 13um of infrared radiation Based on this principle according to the relationship between surface forehead temperature and human forehead s temperature we can measure the human forehead s actual temperature through measuring surface forehead temperature 2 Safety precautions Warning Use of this thermomete...

Page 3: ...efore measurement To make the measurement data reliable and stable when ambient temperature varies a lot the thermometer should be placed indoors for 30 minutes before using Do not measure the sites of scarred tissue or tissue compromised by skin disorders Do not measure the site of forehead temperature if patient has trauma on forehead Do not measure if a patient is treated with certain drug ther...

Page 4: ...racy of measurement data please don t take measurement of body temperature in strong electro magnetic interference environment such as microwave high frequency equipment operation environment Do not disassemble repair or modify the unit Don t touch the battery output when measuring The thermometer must be stored according to the technical specifications The materials ABS of expected contact with th...

Page 5: ...trical thermometer 2 Object temperature measurement mode you can measure the surface temperature of object such as bathwater and milk temperature etc Normal temperature Normal forehead temperature range for different ages Measuring position Normal temperature Normal temperature Anus 36 6 38 0 97 9 100 4 Oral 35 5 37 5 95 9 99 5 Armpit 34 7 37 3 94 5 99 1 Forehead 35 8 38 0 96 4 100 4 Ages Normal te...

Page 6: ...ic data hold and auto power save Automatic selection range and display resolution 0 1 0 1 3 colors backlight 2 measurement mode Forehead and surface mode 1 IR sensor 3 Mode selection 5 UP button 7 Measurement Trigger 2 LCD display 4 Down button 6 Mode button 8 Battery cover 6 Overall description 5 ...

Page 7: ...please stop using 3 Check thermometer When you press the Power Scan button the system will have self testing of software and hardware If there are problems LCD will display Err symbol Check if the sensor laser is dirty damaged 4 Accuracy of unexpected fluctuations in ambient temperature may decrease the measurement results 5 If you want measure forehead temperature clean forehead and arrange hair m...

Page 8: ...should adjust the temperature on the device and add the difference i e 1 1 2 2 To do it press the MODE button for 2 seconds the screen displays F1 press MODE button again until you get F3 press UP button in order to add the difference in our example 1 1 2 2 3rd step To check take the temperature again using the product 2 For the first use or when inserting new batteries please wait about 10 minutes...

Page 9: ...ead or cosmetic can affect the accuracy of measurement please maintain the cleanness of the forehead when measuring Don t use the thermometer outdoors 6 Select object mode for measuring room temperature surface temperature etc TEMPERATURE TAKING HINTS For heat insulated object please don t measure the surface For hot liquid please do not directly measure the surface because the hot steam fog can c...

Page 10: ...1 DOWN to reduce it by 0 1 0 1 The alarm threshold default value is 38 0 100 4 3 F3 Set difference Press MODE button for 2 seconds the screen display F1 press two times to get F3 select UP to increase the difference by 0 1 0 1 DOWN to reduce it by 0 1 0 1 In the case of seasonal or environmental changes a verification and adjustment should be carried out Note This function is only effective body 4...

Page 11: ...ut the mode selection button on the right measurement range 0 to 100 32 0 to 212 0 6 3 colors backlight In body mode the device is able to judge whether a fever or not by different color of backlight 32 0 to 37 4 86 0 to 99 3 Green 37 5 to 37 9 99 5 to 100 2 Orange 38 0 to 43 0 100 4 to 109 4 Red In surface mode the backlight always keeps green 7 Data memory Date memory automatically after tempera...

Page 12: ...ment exceed the 10 40 range will be showing Err 2 If the measured temperature result is 32 0 the LCD will be showing Lo 3 If the measured temperature result is 43 0 the LCD will be showing Hi 4 If the measured temperature result is 38 0 the buzzer will be send out Beep Beep Beep a long one and two short ones 5 We recommend measurement for 3 times to reference and keep the interval between the two ...

Page 13: ...n restore up to 99 sets data the query interface can display memory serial number body mode and recorded tempera ture values Remark The temperature result of surface mode can not be saved And the abnormal body temperature result can not be saved too such as Lo Err or Hi 7 Backlight function The backlight will be according to the different measuring temperature to lighting different color The detaile...

Page 14: ...o restore the factory default status The factory default setup is and Body measurement mode 9 Battery installment and replacement 1 When the quantity of electric charge is not sufficient the icon of low battery will appear on the lower part of the screen Though the battery still can be used it would be perfect to replace it 2 When the quantity of electric charge of the battery is lower than the mini...

Page 15: ...cessories Please contact the retailer for repair 2 Opening of the equipment by unauthorized agencies is not allowed and will terminate any claim to warranty product don t drop the thermometer in water or other liquid 9 Maintenance WARNING No modification of this equipment is allowed 10 Calibration The thermometer is initially calibrated at the time of manufacture If this thermometer is used accordi...

Page 16: ...r conditioner Lo Less than In the object measurement mode The measuring distance is too far Have water vapor condenses on the lens Lo Less than The ambient temperature is beyond of range of measurement The sensor or hardware is damaged Lower battery however you can t use it Replace the new battery Keep the thermometer in the room Whose Temperature is for minutes Excluding the possibility of temper...

Page 17: ...hock Applied part Mode of operation Measuring range Measuring accuracy At laboratory conditions Infrared Thermometer KRT01 Forehead and object temperature modes d c V pcs AAA battery NOT INCLUDED Within Relative humidity KPa Relative humidity KPa L W H mm Approx g including battery IP Internally powered ME equipment Type BF applied part including the whole unit Continuous operation For forehead te...

Page 18: ...asic safety and essential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in home healthcare environment Clinical thermometers Part Performance of infra red ear thermometers with maximum device Medical devices Application of usability engineering to medical devices Biological evaluation of medical devices Part Evaluation and testing...

Page 19: ...ean Community The European representative is Shanghai International Holding Corp GmbH Europe IP code of the device this device s grade of against ingress of solid foreign objects mm diameter and the against access to hazardous parts with finger the grade of waterproof is dripping tilted Manufacturer information The manufacturer is Kross Direct IP 16 Disposal 17 Normalized symbols If you have any qu...

Page 20: ...tions walkie talkies can effect this equipment and should be kept at least a distance d 3 3m away from the equipment Note As indicated in Table 6 of IEC 60601 1 2 2007 of ME EQUIPMENT a typical cell phone with a maximum output power of 2W yields d 3 3m at an IMMUNITY LEVEL of 3V m 19 96 100 Carrier Drive Etobicoke ON M9W 5R1 www krossdirect ca ...

Page 21: ......

Page 22: ......

Reviews: