SECTION 5
REMOTE OPERATION
99
5.7
REMOTE CONTROL EXAMPLES
The REN (remote enable) line on the bus interface must be kept true to maintain REMOTE CONTROL.
REMOTE CONTROL can only be initiated while the UPC is displaying a Metering display or while in the
UPC STATUS display, not while editing.
Be sure that the UPC GPIB device address is set to a known unique address within the GPIB system.
The factory default device address is 1. See section 4.6.3.3 for configuration of the GPIB device
address.
See section 10 for setting up the RS-232 SERIAL INTERFACE parameters if this option is used.
Optional keywords have been omitted throughout this section for clarity. The command strings may
contain spaces.
NOTE:
Un-address means to send the UNTALK and UNLISTEN messages. This is only applicable for
the GPIB, NOT the RS-232 option. With many programming languages that support GPIB
communications, listener and talker addressing and un-addressing is handled automatically and is
transparent to the user. Check your programming language or driver documentation for verification.
NOTE:
<eos> should be a LINE FEED (0Ahex, 10dec). See GPIB convention #16 (Section 5.2) for
more details.
CAUTION
-
ALL :SOURce PARAMETERS SHOULD BE SET
PRIOR TO ENABLING THE OUTPUT FOR THE FIRST TIME
AFTER POWER-ON OR AFTER RESETTING THE UNIT.
THE PARAMETERS MAY BE SET INDIVIDUALLY OR
BY USING :DEFine AND :EXECute OF A STORED PROGRAM
SEE PARA. 5.7.1, 5.7.3 AND 5.7.5
!
Summary of Contents for UPC-32
Page 2: ......
Page 12: ...SECTION 1 GENERAL viii THIS PAGE INTENTIONALLY BLANK...
Page 17: ...SECTION 1 GENERAL 5 PACIFIC Figure 1 3a SCU UPC 32 Figure 1 3b UPC 32 Figure 1 3c UPC 12...
Page 46: ...SECTION 4 OPERATION 34 PA CIFIC FIGURE 4 1 POWER SOURCE FRONT PANEL...
Page 77: ...SECTION 4 OPERATION 65 90 0 100 0 85 95 100 90 FIGURE 4 6 2 1 EDITED WAVEFORM...
Page 139: ...NOTES 127 NOTES...