91
SECTION 6
CALIBRATION
6.0 CALIBRATION
This section describes calibration of the AMX-Series AC Power Source.
6.1 CALIBRATION
INTERVAL
Each AMX-Series Power Source requires calibration once every six months or after service
has been performed to the system.
6.2
TEST EQUIPMENT REQUIREMENTS
The test equipment listed below is required for calibration of the AMX-Series Power Source.
1.
Digital Voltmeter:
4½ Digit True-RMS responding
5000 Hz bandwidth, min.
2.
Frequency counter:
5 digit counter, min.
3.
Digital Ammeter:
3½ Digit True-RMS responding
5000 Hz bandwidth, min.
(Alternate Approach: Current transformer used in conjunction with the DVM.)
4. Oscilloscope:
(Optional)
5.
Load:
Varies by model number;
2A @ 125 VAC, Model 305-AMX, 308-AMX
2.6A
@ 150 VAC, Model 312-AMX
3A @ 125 VAC, Model 105-AMX, 108-AMX
4A @ 150 VAC, Model 112-AMX
6A @ 125 VAC, Model 320-AMX
10A @ 125 VAC, Model 125-AMX
12A @ 125 VAC, Model 345-AMX
16A @ 125 VAC, Model 140-AMX
16A @ 125 VAC, Model 160-AMX
16A @ 125 VAC, Model 360-AMX
24A @ 125 VAC, Model 390-AMX
32A @ 125 VAC, Model 3120-AMX