SPECIFICATIONS
No.
Item
Rating
1.
Delivery Rate (Lower delivery rates available for
low application)
2 to10 LPM(OX-10A)
0.5 to 5 LPM(OX-5A)
2.
Outlet Pressure
8.5-13 psi ( 68.5-90 kpa )
3.
Electrical Rating
230 V, 50 Hz, 3.4A (Max.)OX-10A
1.4A (Max.)OX-5A
4.
Oxygen Percentage
2~9 LPM 93%±3%,10 LPM 90%±3%(OX-10A)
0.5~5 LPM 93%±3%,(OX-5A)
5.
Average oxygen concentration
at 2 l/min: 93%±3%
6.
Operating Environment Range
Temperature 10
℃
to 40
℃
,
Humidity 15% to 95%
Atmospheric pressure 50~106 kpa (7.3~15.4 psi)
7.
Weight
18.9kg(OX-10A)/ 14.9kg(OX-5A)
8.
Sound Level ( ISO 8359:1996 from front )
≤
55dbA (OX-10A) /
≤
43dbA (OX-5A)
9.
Dimensions
342mm(L) x368mm(W) x572mm(H)
10.
Operating System
Time Cycle / Pressure Swing
11.
Storage Conditions
Temperature -40
℃
to 70
℃
Humidity 10% t0 100%
Atmospheric pressure 50~106 kpa (7.3~15.4 psi)
12.
Equipment Class and Type
The type of protect against electric shock: Class II
The degree of protect against electric shock: TYPE B Type B
Applied part
13.
Pressure Relief Valve
276 kPa±34.5 kPa
14.
Lifetime
3 Years
15.
Warranty Period limit
1 Year ( OX-10A) / 3 Years ( OX-5A)
Page 13 of 16