USER’S MANUAL FOR THE DEVICE NOCOSPRAY (230V) 

 
 
 

I – PRESENTATION OF THE DEVICE 

 
 
 

 
  1 – Power Cable 
  2 – Used fuse 
  3 – Switch for power On/Off 
  4 – Handle 
  5 – Button of nose of the diffuser 
  6 – Nose of the diffuser 
  7 – Bottle (1 litre) 
  8 – Button to activate the spray, “Start“ button 
  9 – Jet 
10 – Filtering foam 
11 – Selector for  volume to be dealt with 
12 – USB tracking (option) 
13 – Switch for timer (option) 
 
 

CHARACTERISTICS 

 

Power 1000 

Tension 

230 V mono+T 

Frequency 

50 to 60 Hz 

Intensity 4,5 

Maximum Intensity  

 

10 A 

Speed of rotation of the turbine 

22000 rpm 

Net weight (without liquid or bottle)   5,8 kg 
 

PERFORMANCES 

 
Exit speed of the air 

 

80 m/s 

Average output of the liquid 

1000 ml/h 

 
 
 

 

 

 

 

GENERAL INFORMATION: 
 

   

- Type: Diffuser 
- Name: Nocospray 
- Humidity of functioning and storage 25% to 75% 
- Do not use in presence of a flammable anaesthetic mixture 

- Protection factor: IP10. 
- Storage temperature 0°C to 50°C.

 

- Electric classification: Class 1 
- Intermittent service

.

 

 

II –  PRECAUTIONS FOR USE  

 

 

The manufacturer will not be held liable for any bodily harm or material damages caused by inappropriate or abnormal use of the 
devise contrary to the precautions recommended in the list below (non exhaustive list): 
 - Use only with the products of the range Oxy'Pharm 
 - Make sure that electric power matches regular standards 
 - Do not introduce an object or liquid in the nose of NOCOSPRAY while it is working or not working 
 - Avoid moving NOCOSPRAY when it contains of the product, or at least, to maintain him (it) very horizontally 
 - Check the presence of the jet (

9) 

 - Never fill up the bottles otherwise than with the product of the OXY'PHARM

® 

range 

 - Remove the bottle of the NOCOSPRAY for transportation 
 
The device diffuses a chemical substance; therefore it is necessary to pay attention: 
 - Not to inhale the product 
 - 

Not to smoke near the NOCOSPRAY while it is working

 

 - Not to steer it towards an incandescent zone 
 - Not to be located in the room during the spray 
 

Summary of Contents for NOCOSPRAY

Page 1: ...n lectrique Classe 1 Ne pas utiliser en pr sence d un m lange anesth sique inflammable Service intermittent SYMBOLE DE SECURITE consulter la notice terre de protection II PRECAUTIONS D EMPLOI La respo...

Page 2: ...suffit de basculer l interrupteur pour programmateur 13 sur la position avec programmateur Sur cette position l appareil d marre directement d s qu il sera branch et que l interrupteur g n ral 3 est a...

Page 3: ...ont axe En cas de mauvaise compr hension ou de doute sur la lecture du guide d utilisation et d entretien veuillez nous contacter au 33 1 48 82 22 22 Distribu par OXY PHARM Fabriqu par AIREL ZA des Gr...

Page 4: ...n cas d intervention et r paration par des personnes trang res la S A R L AIREL Les scell s de l appareil doivent tre intactes Par ailleurs si une corrosion anormale des parties souples durite joints...

Page 5: ...rage temperature 0 C to 50 C Electric classification Class 1 Intermittent service II PRECAUTIONS FOR USE The manufacturer will not be held liable for any bodily harm or material damages caused by inap...

Page 6: ...WITH TIMER OPTION To use the appliance with a timer change the position of the Switch for timer 13 on position with timer On this position the appliance will start directly when plugged in and the ge...

Page 7: ...ing or doubts reading this User s Maintenance Manual please contact us at 331 48 82 22 22 Distributed by OXY PHARM Produced by AIREL ZA des Grands Godets 917 rue Marcel Paul 94508 Champigny sur Marne...

Page 8: ...of intervention and repair by persons not member of the AIREL Company The seals of the device have to be untouched Besides if an abnormal corrosion of supple parts radiator hoses joints was noticed in...

Reviews: