33
Cleaning, Care and Maintenance Chapter 6
English
4. Install a replacement fuse, Inogen RP#125 (BUSS MDA-12) and reassemble the tip.
Ensure the retainer ring is properly seated and tightened.
Inogen One® G4 Column Change Procedure
NOTE
Column change instructions are only to be used when maintenance is
required and are not intended for practice purpose.
1. Turn off the Inogen One® G4 concentrator by pressing the power button to shut
down the device.
2. Remove the Inogen One® G4 concentrator from the carry case, if applicable.
3. Remove the battery from the Inogen One® G4 concentrator.
4. Place the Inogen One® G4 concentrator on its side so
that the underside is visible. The metal column as-
sembly can be seen on one side of the device.
Fuse
Cigarette Adapter Plug
Retainer Tip
Retainer
CAUTION
For continued protection against risk of fire, only use fuse specified.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019