Oxidation Technologies OA-1 Installation And Operation Manual Download Page 6

Operation

- Plug OA-1 into 120 VAC outlet.

- The alarm will engage immediately upon powering it up -  This is because the sensor 
needs to warm up. Initial warm up may take from 30 minutes to 1 hour - 12 hours for 
full warm up. 

- After 30 minutes of warm up, press the brown button on the circuit board inside to 
perform a zero calibration, make sure this is done in an ozone free environment.

- After 12 hours of warm up, perform a zero calibration once more.

-  Set ozone alarm set point – The dial in the upper right corner is 100% of sensor range.

Example - If sensor range is 1 ppm, and the alarm is to be set at .1 ppm, the dial will be 
set to 1.

Page 6

214 West Highway 18

Inwood, IA 51240

PH: (515) 635-5854

Summary of Contents for OA-1

Page 1: ...OA 1 Ozone Alarm Installation and Operation Manual Page 1 214 West Highway 18 Inwood IA 51240 PH 515 635 5854...

Page 2: ...ements concerning the use of ozone WARNING Ozone is an extremely aggressive and powerful oxidizer The Occupational Safety and Health Administration OSHA 8 hour exposure limit is 0 10 PPM The OSHA 15 m...

Page 3: ...arnings and Hazards 2 Table of Contents 3 Introduction 4 Features 4 Specifications 5 Installation 5 Operation 6 Internal Components 7 Wiring Diagram 8 System Maintenance 9 Contact Info 10 Page 3 214 W...

Page 4: ...tures Red strobe light will alert of ozone levels above set point Audible siren will alert of ozone levels above set point Volume can be adjusted simply by turning the cover of the siren User settable...

Page 5: ...e semiconductor sensor Size 11 long x 4 5 deep x 8 wide Weight 1 lb Power requirements 12 VDC at 300 mA AC adapters available worldwide Relay Outputs 10 amp relay Analog output 4 20 mA output linear w...

Page 6: ...m up press the brown button on the circuit board inside to perform a zero calibration make sure this is done in an ozone free environment After 12 hours of warm up perform a zero calibration once more...

Page 7: ...Internal components Page 7 214 West Highway 18 Inwood IA 51240 PH 515 635 5854...

Page 8: ...Wiring Diagram Page 8 214 West Highway 18 Inwood IA 51240 PH 515 635 5854...

Page 9: ...of any of these components fail please contact the manufacturer MONTHLY Perform a zero calibration Clean Alarm Bump test with ozone ANNUALLY SM 3 Ozone Leak Sensor Replace Page 9 214 West Highway 18...

Page 10: ...ontact Info By mail Oxidation Technologies LLC 214 West Highway 18 Inwood IA 51240 By telephone 515 635 5854 By e mail info oxidationtech com Page 10 214 West Highway 18 Inwood IA 51240 PH 515 635 585...

Reviews: