Oxidation Technologies OA-1 Installation And Operation Manual Download Page 5

Specifications

Range: 

Three ranges available

- 0-0.1 ppm
- 0-1.0 ppm
- 0-10.0 ppm

Resolution: 

Range dependent

- 0.1 ppm range = 0.01 ppm resolution
- 1.0 ppm range = 0.01 ppm resolution
- 10.0 ppm range = 0.1 ppm resolution

Graphic Display: 

Indicates the current ozone level and past 12 hour trend graph

Response time: 

Within 8 seconds

Measurement principle: 

HMOS (heated metal oxide semiconductor) sensor

Size: 

11" long x 4.5" deep x 8" wide

Weight: 

1 lb

Power requirements:

12 VDC at 300 mA. AC adapters available worldwide.

Relay Outputs:

10 amp relay

Analog output: 

4-20 mA output linear with ozone level

Calibration: 

Replace your sensor annually

Installation

Mount the OA-1 Ozone Alarm in a clean, dust free environment that is non condensing.
Recommended mounting height is 5-6 ft.
Ambient Temperature should be between 50 and 86 deg. F.
Ambient Humidity should be in the range of 25 to 75% RH

Page 5

214 West Highway 18

Inwood, IA 51240

PH: (515) 635-5854

Summary of Contents for OA-1

Page 1: ...OA 1 Ozone Alarm Installation and Operation Manual Page 1 214 West Highway 18 Inwood IA 51240 PH 515 635 5854...

Page 2: ...ements concerning the use of ozone WARNING Ozone is an extremely aggressive and powerful oxidizer The Occupational Safety and Health Administration OSHA 8 hour exposure limit is 0 10 PPM The OSHA 15 m...

Page 3: ...arnings and Hazards 2 Table of Contents 3 Introduction 4 Features 4 Specifications 5 Installation 5 Operation 6 Internal Components 7 Wiring Diagram 8 System Maintenance 9 Contact Info 10 Page 3 214 W...

Page 4: ...tures Red strobe light will alert of ozone levels above set point Audible siren will alert of ozone levels above set point Volume can be adjusted simply by turning the cover of the siren User settable...

Page 5: ...e semiconductor sensor Size 11 long x 4 5 deep x 8 wide Weight 1 lb Power requirements 12 VDC at 300 mA AC adapters available worldwide Relay Outputs 10 amp relay Analog output 4 20 mA output linear w...

Page 6: ...m up press the brown button on the circuit board inside to perform a zero calibration make sure this is done in an ozone free environment After 12 hours of warm up perform a zero calibration once more...

Page 7: ...Internal components Page 7 214 West Highway 18 Inwood IA 51240 PH 515 635 5854...

Page 8: ...Wiring Diagram Page 8 214 West Highway 18 Inwood IA 51240 PH 515 635 5854...

Page 9: ...of any of these components fail please contact the manufacturer MONTHLY Perform a zero calibration Clean Alarm Bump test with ozone ANNUALLY SM 3 Ozone Leak Sensor Replace Page 9 214 West Highway 18...

Page 10: ...ontact Info By mail Oxidation Technologies LLC 214 West Highway 18 Inwood IA 51240 By telephone 515 635 5854 By e mail info oxidationtech com Page 10 214 West Highway 18 Inwood IA 51240 PH 515 635 585...

Reviews: