Specifications
Range:
Three ranges available
- 0-0.1 ppm
- 0-1.0 ppm
- 0-10.0 ppm
Resolution:
Range dependent
- 0.1 ppm range = 0.01 ppm resolution
- 1.0 ppm range = 0.01 ppm resolution
- 10.0 ppm range = 0.1 ppm resolution
Graphic Display:
Indicates the current ozone level and past 12 hour trend graph
Response time:
Within 8 seconds
Measurement principle:
HMOS (heated metal oxide semiconductor) sensor
Size:
11" long x 4.5" deep x 8" wide
Weight:
1 lb
Power requirements:
12 VDC at 300 mA. AC adapters available worldwide.
Relay Outputs:
10 amp relay
Analog output:
4-20 mA output linear with ozone level
Calibration:
Replace your sensor annually
Installation
Mount the OA-1 Ozone Alarm in a clean, dust free environment that is non condensing.
Recommended mounting height is 5-6 ft.
Ambient Temperature should be between 50 and 86 deg. F.
Ambient Humidity should be in the range of 25 to 75% RH