3. SAFETY PRECAUTIONS
The personal micro centrifuge is designed to be safe under the
following conditions when operated with necessary precautions
mentioned below :
• Indoor use only.
• Ambient temperature of 4
°
C to 6S
0
C.
• Main supply voltage fluctuations not to /-10% of the
nominal voltage.
• Position at least 20 cm from walls, 20 cm from ceiling and 20 cm
between other products.
• Make sure to always install the centrifuge on stable flat surfaces only
& place it near the electrical outlet.
• Always load the rotor symmetrically. Each tube should be
counterbalanced by another tube.
• Make sure to always use only specified micro centrifuge tubes made
from plastic and designed to withstand centrifugal forces of at least
2000g.
• Do not operate the centrifuge without the rotor being properly
attached to the shaft. Re-check by trying to pull it up manually,
especially after the rotor has been changed or re-assembled.
• Do not move the centrifuge while it's under operation.
• Do not fill the tubes while they are in the rotor. Liquid spillage may
harm the unit.
• Do not put hands in the rotor chamber unless the rotor has
completely stopped.
• Do not centrifuge flammable, explosive or corrosive materials.