Battery Replacement

Remove the battery compartment lid and replace one SR44 with its po) side facing upward.

Batteries

Order Code

SR44B Silver Oxide Battery (Pack 1)

PAN-904-5020K

Troubleshooting

Failure

Five digits appear simultaneously and change rapidly

Possible Cause

Battery Power Low

Solutions

Remove battery compartment lid and replace battery

Failure

The display does not work after changing battery

Possible Cause

Circuit accidentally broken by poor voltage when changing battery

Solutions

Remove battery and replace it within 30 seconds

Failure

Reading error for full length is 

0.1mm

Possible Cause

Dirt has damaged/covered sensor

Solutions

Remove cover and slider assembly. Apply a small layer of petrol to avoid dirt build-up

Failure

Nothing appears on LCD

Possible Cause

Battery has poor contact. Battery voltage is under 1.1v

Solutions

Check contact of battery or replace the battery