Battery Replacement
Remove the battery compartment lid and replace one SR44 with its po) side facing upward.
Batteries
Order Code
SR44B Silver Oxide Battery (Pack 1)
PAN-904-5020K
Troubleshooting
Failure
Five digits appear simultaneously and change rapidly
Possible Cause
Battery Power Low
Solutions
Remove battery compartment lid and replace battery
Failure
The display does not work after changing battery
Possible Cause
Circuit accidentally broken by poor voltage when changing battery
Solutions
Remove battery and replace it within 30 seconds
Failure
Reading error for full length is
≤
0.1mm
Possible Cause
Dirt has damaged/covered sensor
Solutions
Remove cover and slider assembly. Apply a small layer of petrol to avoid dirt build-up
Failure
Nothing appears on LCD
Possible Cause
Battery has poor contact. Battery voltage is under 1.1v
Solutions
Check contact of battery or replace the battery