Version 1.4 rev 24 Apr 2019
25
IKon-M
5.1.3 M
oUnting
tHe
c
aMeRa
on
tHe
v
acUUM
c
HaMbeR
f
lange
1. The SO version of the iKon-M is fitted with a 6” ICF flange as standard with 16 holes. The holes are on a
rotatable rim which allows adjustment to match the hole pattern on the customer’s vacuum chamber with either
a hole indexed to the Top Dead Centre (TDC) position or with holes midway, (360/16)/2 = 11.25° either side of
TDC.
2. A ball spring plunger in the circumference of the rotatable rim will index to positions on, or either side of TDC
with reference to the space between water connectors.
3. Up to 6 of the 16 holes may be obstructed from screw access depending on the orientation of the rim. These
positions will be behind the water connectors and at the bottom corners of the camera body, i.e.:
4. The rotatable rim allows these holes to be moved clear of the obstructions so the screws can be loaded. The
ball spring plunger may need to be loosened slightly to allow rotation.