Version 1.4 rev 24 Apr 2019
15
IKon-M
2.2 R
eaR
p
anel
Figure 3: Rear Panel
USB 2.0
Connects the iKon-M to the control PC via a USB 2.0 connection.
TTL / Logic
The TTL/Logic connection permit connection to other devices for synchronisation and control of fire, trigger and shutter
operations. Connector type: SMA, provided with SMA - BNC cable Fire (Output), External Trigger (Input), Shutter
(Output). The SMB outputs (Fire & Shutter) are CMOS compatible & Series terminated at source (i.e. in the camera
head) for 50Ω cable. NOTE: The termination at the customer end should be high impedance (not 50Ω) as an incorrect
impedance match could cause errors with timing and triggering. The SMB Ext Trig input is TTL level & CMOS compatible
and has 470Ω impedance.
I
2
C
The I
2
C provides a connection option via the I
2
C interface. Compatible with Fischer type SC102A054-130 connector. (1)
Shutter (TTL), (2) I
2
C Clock, (3) I
2
C Data, (4) +5 Vdc, (5) Ground.
Power
Power input connection for connection to the PSU. Refer to
Section 2.3.
Notes:
Minimum cable clearance required at rear of camera 90 mm.
Liquid cooling connections provide the facility for connection to re-circulator or other water/liquid cooling system. Refer
to
Section 3.7
.
I
2
C
1
2
3
4
5
TTL/Logic
USB 2.0
Power Input
Water/Liquid cooling
connectors (x2)