Version 1.4 rev 23 Apr 2019
35
IKon-L
APPENDIX A: TECHNICAL SPECIFICATIONS
(Note information given for iKon-L 936, refer to specific product specification sheet for model variations)
General Specifications
Weight
4.6 kg [10 lb 2 oz], PS-29 = 0.7 kg [1 lb 6 oz] approx
Dimensions mm [inches]
W x H x D
128.0 x 153.7 x 209.5 mm [5.04 x 6.05 x 8.29 inches]
Environmental Specifications
Usage
Indoor use only
Operating Temperature
0°C to 30 °C ambient
Storage Temperature
-25°C to 50 °C
Operating Relative Humidity
<70% (non-condensing)
Ingress Protection
IP20
Overvoltage Category
CAT II. An overvoltage category of CAT II means that the equipment is designed to cope
with transient voltages above the rated supply that would be experienced by any product
connected to a mains socket in a building.
Rated Pollution
Pollution Degree 2. Normally only non-conductive pollution occurs. Occasionally, however, a
temporary conductivity caused by condensation must be expected.
Cooling vent clearance
100 mm minimum
Power Requirements
Mains Input
100-240 V AC 50/60 Hz
Cable pin-outs
(see Section 2.3)
Power Consumption
iKon-L 936 with 4 stage cooler (W) 60W max
iKon-L 936 with 5 stage cooler (Z) 120W max
iKon-L HF with 2 stage cooler (F) – 142W Max
Model Specific Information
Refer to the specification sheet for further technical and performance information required model:
•
•
•
•