Version 1.4 rev 23 Apr 2019
17
IKon-L
2.3 p
oWeR
s
Upply
U
nit
and
p
oWeR
R
eqUiReMents
The iKon-L uses a dual PSU setup comprised of a PS-40 and a PS-29. The PS-29 (or equivalent user supplied unit) is
connected to the mains electrical supply via an IEC lead and performs AC-DC conversion. The PS-40 takes the 24 VDC
output of the PS-29 and supplies the various DC voltages required by the camera. The PS-40 is connected to the iKon-L
via a 16 pin cable. Connection of the power supplies during installation is outlined in
Section 3.6
.
PS-29
Input
: 100-240 VAC, 47-63 Hz, 2.2 A max. Cable input is IEC, connection at mains to match regional requirements.
Output
: 24 VDC @ 150 Watts/6.6 A max, 4 pin cable (see below).
Pin
Description
1
24 VDC (+ve)
2
24 VDC (+ve)
3
(-ve)
4
(-ve)
Shell
GND
PS-40
Input:
24 VDC, 13 A max. Power on/off switch on input panel.
Output
: See below
Aux Output:
24 VDC 1A max. May be used to power other devices.
Pin
ID
Pin
ID
Pin
ID
Pin
ID
1
+8V
5
GND
9
COOLER-
13
-70V
2
-8V
6
10
+5V
14
DIFF_GND
3
-15V
7
11
GND
15
DIAL-A-TEMP
4
+15V
8
COOLER-
12
Spare 1
16
+54V
Notes:
•
If users use any other power supply they do so at their own risk.
•
Do not obstruct ventilation holes. Unit may become hot during operation.
PS-29
PS-40
iKon-L
From Mains supply
24 VDC
4-pin
Cable
16-pin
Cable