TROUBLE SHOOTING
KMW200CCD
Version 1.1
USER MANUAL
KMW200CCD Chiller v1.1
Page
20
9. Trouble Shooting
Symptom Fault
Solution
Pump stops after about
10 seconds and the red
LEVEL light is
illuminated
Water level in the reservoir is too low Fill the reservoir with water
(refer to the MAINTENANCE
INSTRUCTIONS chapter)
Pump stops after about
10 seconds and the red
FLOW light is
illuminated
Rate of water flow is too low
Clean the water filter (refer to
the MAINTENANCE
INSTRUCTIONS chapter)
Red TEMP light
illuminated and
KMW200CCD stops
working after 10
minutes
Temperature of the water in the
water reservoir exceeds 18
°
C
Cool the water in the water
reservoir
Fuses
Designators Value
Package
Location
F1
315mA / 250V
5x20mm
Inside KMW200CCD (PCB)
F2
6.3A / 250V
5x20mm
Inside KMW200CCD (PCB)