TECHNICAL DESCRIPTION
KMW200CCD
Version 1.1
USER MANUAL
KMW200CCD Chiller v1.1
Page
8
2. The level of the water in the reservoir
3. The temperature of the water in the water reservoir
The pump stops after 10 seconds if either the rate of water flow, or the water level in the reservoir,
is too low.
The pump stops after 10 minutes if the water temperature in the water reservoir exceeds 18
°
C.
The state of the unit is displayed on four status lights on the front panel.
Figure 4.2.2 Water Conditioning Check Routine of the KMW200CCD
READY Lamp ON
CCD output ON
Key switch ON
Pump ON
Flow
Level
Temperature
READY Lamp OFF
CCD output OFF
Sensor’s
subroutine
TRUE
FALSE
LOOP FOR EVER
Wait 10 sec
PUMP OFF