HANDLING, INSTALLATION, STORAGE AND TRANSIT INFORMATION
KMW200CCD
Version 1.1
USER MANUAL
KMW200CCD Chiller v1.1
Page
14
Water fill cap
Figure 5.2.3 Top view of the KMW200CCD showing the
water
refill cap
3. Place a bucket under the overflow pipe at the rear of the KMW200CCD.
WARNING
Do not allow water to splash on the electrical circuits of the
KMW200CCD. Failure to do this could cause an electrical shock
hazard, and possible damage to the equipment.
4. Using the funnel, fill the reservoir with water, until water starts to come out of the overflow pipe.
5. Refit the water fill cover on the KMW200CCD.
6. Open the manual valves on the side of the KMW200CCD.
CAUTION
When inserting the CCD detector power supply cable into the KMW200CCD, ensure
that the connectors are pushed home firmly.
7. Connect the power cable for the CCD detector between the connector on the CCD detector
power supply and the socket marked as CCD on the KMW200CCD. Refer to figure 5.2.4.
Figure 5.2.4 Electrical power supply to the KMW200CCD
8. Plug the KMW200CCD power cable into the socket marked
Mains IN
on the KMW200CCD.
9. Plug the other end of the KMW200CCD power cable into the mains electrical supply.
10. The KMW200CCD is now ready to use.
Mains
supply
KMW200CCD
CCD detector’s power
supply unit
Mains IN
CCD