background image

E-935-1-E

3

1. GENERAL

This pulse checker is designed for inspecting, adjusting and maintaining a wide variety of equipment 

associated with pulse signals. It features two groups of functions - the input functions which include the 

accumulation, frequency measurement, period measurement and gated accumulation of input pulses and 

the output functions which include the digital selection of the reference frequency, frequency reduction 

and output frequency programming by selecting % frequency reduction, the control of the number of 

output pulses, as well as the selection of pulse width and generator type.

Despite these elaborate capabilities, this portable pulse checker is easy to use.

To derive maximum bene

t from the instrument, we recommend that you familiarize yourself with the 

instructions contained in this manual.

2.  PART NAMES

No.

NAME

DESCRIPTION

1

DC IN

Connector to the ac adaptor furnished

2

POWER

Power switch

3

FUNCTION

Input counter function select rotary switch

4

INPUT LEVEL

Input pulse trigger level select rotary switch

5

INPUT COUNTER

8-digit LCD Input counter

6

RESET

Input counter reset pushbutton switch

7

INPUT

Input terminals to banana plugs of the furnished cords

8

FREQUENCY

4-digit decade switch for establishing the reference frequency

9

PRESET

8-digit decade switch for establishing the output pulse number

10

START

Output counter "Start" pushbutton switch, LED (green) lighted

11

RESET

Output counter "Reset" pushbutton switch, LED (yellow) lighted

12

STOP

Output counter "Stop" pushbutton switch, LED (red) lighted

13

OUTPUT COUNTER

8-digit LCD output counter

14

RATING

Reference frequency reduction rotary switch

15

%SETTER

Reference frequency % setting rotary switch

16

WIDTH

Output pulse width setting rotary switch

17

OUTPUT

Output generator type select rotary switch

18

OUTPUT

Output terminals to banana plugs of the furnished cords

19

Battery compartment cover

20

Carrying strap attachment 

tting

21

Rubber foot

SUP

SIG

SIG

GATE

OV

OV

OUTPUT

INPUT

INPUT LEVEL

INPUT COUNTER

OUTPUT COUNTER

FUNCTION

OUTPUT              WIDTH         %SETTER        RATING         STOP  RESET  START

PULSE  CHECKER

PC  2201

PRESET

FREQUENCY RESET

POWER

D

C

 IN

100

10

1

1K

10K

30

75

100

25

0

1

50

0.25

1.1

(ma)

PG30S

O. C

VOLT
PA11

P. A

PG30

PG20

OPTO

PERIOD

RATIO

105

15

COUNT

FREO

3

6

15

D.C

1.5

(V)

18

5

7

8

6

9

20

21

1

2

19

10

11

12

13

14

15

16

17

3

4

Summary of Contents for PC2201

Page 1: ...SE CHECKER This product is fabricated and shipped from our factory under stringent quality control With the proper application of the information and knowledge contained in this manual you will be abl...

Page 2: ...OTE CAUTION and WARNING as explained below NOTE Notes are separated from the general text to bring the user s attention to important information CAUTION Caution statements inform the user of hazards o...

Page 3: ...t LCD Input counter 6 RESET Input counter reset pushbutton switch 7 INPUT Input terminals to banana plugs of the furnished cords 8 FREQUENCY 4 digit decade switch for establishing the reference freque...

Page 4: ...TION 1 6 SCALER 1 100 END BUZZER BATCH CONTROL CIRCUIT FREQ REDUCTION 8 DIGIT DECADE SWITCH 8 DIGIT LCD DISPLAY COUNT COLLATE OUTPUT COUNTER OUTPUT SIGNAL SELECT CKT PULSE WIDTH SHAPING STABILIZED VOL...

Page 5: ...batteries should be replaced every year although their life may vary depending on the given operating conditions 2 Input Functions 1 INPUT terminals SIG Input terminals to accept pulse signals with se...

Page 6: ...total counter Its functions are selectable with FUNCTION select switch 3 Output Functions 1 FREQUENCY Preset Switch Establishes the reference frequency of output pulses with 4 digit decade switch Pre...

Page 7: ...nhibited while the green COUNTING LED is illuminated however When the accumulated total count on the OUTPUT COUNTER has reached the PRESET count the checker shuts off pulse output generation automatic...

Page 8: ...t of PG30S generator circuit O C A simulated output of FLOWPET NX or PG60A E generator circuit VOLT A voltage pulse output from 8V internal power source PA11 A simulated output of PA11 generator circu...

Page 9: ...0V Vcc 8 5V GND 0V 1k 7 4V 5 6V PC2201 OUTPUT OPTO Vcc 13 5V GND 0V 10k RECOMMENDATION R 10k R 10k 13 4V 0 1V PG20 PG30 Vcc 24V GND 0V 510 22 0V 14 3V P A Vcc 13 5V 30mA GND 0V 10k 6 5V 0 1V Vcc 13 5V...

Page 10: ...GND 0V 10k min 31 9V 1 0V 30mA 0 25 or 1 or 50ms 0 25 or 1 or 50ms PC2201 RECEIVING INSTRUMENT SIG WAVEFORM Voltage levels of signal waveform include an error of 0 5V Signal waveforms shown here indic...

Page 11: ...1 External connection cords terminated with alligator clip and banana plug Red 4 black 2 6 OUTLINE DIMENSIONS 7 STANDARD ACCESSORIES HANDLE SLOT AC ADAPTOR SLOT POWER SWITCH SLOT SOFT CASE 1 EA ACCESS...

Page 12: ...E 935 1 E 2 All specifications are subject to change without notice for improvement 2019 02 Revised E 935 1 E 1...