background image

E--883--4N--E

9

5. GENERAL SPECIFICATIONS

Table 5.1

Item

Description

Input
signals

Flow input

Signal name

Pulse 
generator

Power to generator

Speed of 

response

8V 2-wire current pulse

OPTO 01, 02 8.5VDC

Allowable 

current 

40mA app.

Shortcircuit 

protection 

circuit 

provided

2kHz

Contact closure pulse

PG20

13.5VDC

50Hz

2-wire, 12VDC 3-wire

Voltage pulse 

PG30, 

NPG60A

F

2kHz

24VDC 2-wire

Current pulse (4/20mA)

PA14, 15, 25

NPG60A

E

24.0VDC

12VDC 2-wire

Current pulse

PG30S

13.5VDC

Open collector pulse

FlowPet-5G,

NPG60A (E3)

Temperature input

Resist. thermometer

Pt 100

Ω

 at 0

°

C, 3-wire  Rated current 2mA

Analog

4 to 20mADC or 1 to 5VDC

Output 
signals

Pulse output

Uncorrected total flow

Static relay  Capacity: 230VAC/340VDC 0.2A  

Pulse width: 1ms/50ms

Corrected total flow
Output sync with flow input 

signal

Open collector pulse

Alarm output

Static relay   Capacity: 230VAC/340VDC 0.2A

Analog output

Corrected instant flowrate

4 to 20mADC (Max. load resistance 500

Ω

or 1 to 5VDC (Output impedance 250

Ω

)  

Conversion accuracy: 

±

0.1% of F.S.

Display type

Display: ST display (128

×

128 dots), back lighted                                   (

1), (

2)

Information displayed: Data, units, error message simultaneously

Menu
items

Uncorrected total flow

Same as output pulse unit (m

3

, etc.)

Display capacity: 8 digits

Corrected total flow

Same as output pulse unit (m

3

 (C), etc.)

Uncorr. instant flowrate

m

3

/h, etc.

Corr. instant flowrate

m

3

/h (C), etc.

Temperature

To the 2nd decimal place (when 

°

C is selected.)

Meter error corr. factor To the 5th decimal place
Temp. correction factor To the 5th decimal place
Error alarm

Number of  error messages

Com-

puting 

range

Temperature

Pt 100

Ω

 at 0

Range:

50 to 350

°

C

Std. Span: 70

4 to 20mADC or 1 to 5VDC

Std. Span

200

Com-

puting

accuracy

Corrected total flow

±

0.2% of R.D.

Temperature

Pt 100

Ω

 at 0

±

0.3% of span

4 to 20mADC or 1 to 5VDC

±

0.1% of span

Clock IC battery backup

Lithium battery (Battery life: 10 years approx.)

Communication (when com. 

interface is provided.)

Interface: RS485  Multipoint (Up to 16 units can be connected.)

Dedicated protocol  Baud rate: 4800 bps standard  9600 bps max

Transmission cable

Use 3-conductor shielded cable to the resistance thermometer bulb.

Loop resist. 5

Ω

 max. Example: 300 meters max. with 3-conductor 1.25mm

2

 

cable; 500 meters with 2.0mm

2

 cable

Power source

85 to 264VAC, 50/60Hz or 20 to 30VDC

Power consumption

20W max.

Ambient temperature

10 to +50

°

C

Mounting

Panel mount

Finish

Munsell N1.5

Weight

2.5 kilograms, approx.

NOTES (

1): ST display stands for Super Twisted Nematic display. 

(

2): Backlight life (luminance declined to one half of its original luminance): 2500 hours, approx.

Summary of Contents for EL4131

Page 1: ...nowledge contained in this manual you can expect the best possible results over a long service life of this instrument Keep this manual for ready reference It is suggested that the instruction manuals...

Page 2: ...NTERNAL COMPONENTS AND FUNCTIONS 10 6 1 Front Panel 10 6 1 1 Display 10 6 1 2 On Screen Menu Items 10 6 1 3 Error Messages 10 6 1 4 Front Panel Keypad 11 7 CALCULATION FORMULAS 12 7 1 Implementation o...

Page 3: ...1 170 9 EL4501 175 10 DIP SWITCH CONFIGURATION 182 Model with communication capabilities COMMUNICATION MANUAL Ins No E 880CM E CONTENTS 1 COMMUNICATION CAPABILITY OVERVIEW 5 2 NETWORK CONFIGURATION 5...

Page 4: ...ire supply complete informa tion as to the product name model number product number ratings and other pertinent information 1 2 Transportation Precautions 1 In order to safeguard against damage during...

Page 5: ...the companion flowmeter is simple by keystrokes on the front panel keypad or by inserting an IC card into the slot 2 Built around a microprocessor the instrument carries out calculations entirely in d...

Page 6: ...ed to install instruments alongside each other dimensions in Fig 3 2 are suggested 3 Recommended mounting height is given in Fig 3 3 3 2 3 Installation 1 Front mount the instrument through the cutout...

Page 7: ...s are found on the back of the instrument Fig 4 1 4 3 Description of Terminal Blocks for External Connections Terminal blocks for external connections TB1 TB2 and TB3 are found on the back of the inst...

Page 8: ...4 to 20mADC 1 to 5VDC 5 Temp in b Pt100 3 wire 5 6 B 6 Pulse out 1 No Polarity Uncorrected total flow Static relay 7 A 7 8 4 to 20mADC 1 to 5VDC 8 Pulse out 2 No polarity Corrected total flow Static r...

Page 9: ...Same as output pulse unit m3 etc Display capacity 8 digits Corrected total flow Same as output pulse unit m3 C etc Uncorr instant flowrate m3 h etc Corr instant flowrate m3 h C etc Temperature To the...

Page 10: ...nd corrected for temperature is indicated 5 Temperature TEMPERATURE The temperature value currently fed to the instrument is indicated If a problem arises in the course of temperature conversion such...

Page 11: ...ion keys Four round keys Mainly used for function selection such as activating the input conditions selected and recon figuration Shift keys Four triangular keys Used for moving the cursor moving betw...

Page 12: ...ont panel keys and the frequency pulse information flowrate 7 2 Implementation of Temperature Correction Corrected total flow Q2 Q1 K K a b t0 t c t0 t 2 where Q2 Total flow corrected for temperature...

Page 13: ...ems of the total mass flow and total calorific flow and make sure that incoming pulses are being accumulated Also verify that pulse output and analog output are properly generated 5 Remove the simulat...

Page 14: ...ure correction factor 1 Temperature reading correct 1 A fault in internal assembly While fluid is allowed to flow Fails to count pulses fails to produce a pulse output 1 Input signal line correctly wi...

Page 15: ...e range check 12 A D CONVERT ERROR A D converter circuit error 13 DENSITY CONVERT ERROR Density converter circuit error 14 TEMP 1 PT OVER Temperature Pt input 1 overflow 15 TENP 1 PT UNDER Temperature...

Page 16: ...D Converter Press Input Ckt analog Temp Input Ckt analog Temp Input Ckt Pt Analog Output Circuit Pulse Output Circuit 1 Pulse Output Circuit 2 Pulse Output Circuit 3 Multi plexer Isolation Test Pins...

Page 17: ...e N Non Output item 2 Pulse output 2 B 1 Before correction pulse width 1ms B 5 Before correction pulse width 50ms B 9 Before correction special pulse width C 1 After correction pulse width 1ms C 5 Aft...

Page 18: ...am Japanese D W E Wiring diagram English D A J Compression coef cient calculation process Japanese D A E Compression coef cient calculation process English D T J Inspection procedure Japanese D T E In...

Reviews: