background image

E--883--4N--E

1

11. BEHAVIOR IN ERRATIC CONDITIONS

(1) Erratic A/D Converter:

If, following energization of A/D converter circuit, the conversion process is not completed within 

the specified time period, it is found that the circuit has a problem. At this point, temperature and 

pressure fallback values are used for calculation.

(2) Temperature and Pressure Input Out of Range:

If the input exceeds the high limit or falls below the low limit, its default fallback value is used for 

calculation.

(3) Analog Output Overflow:

A full scale output is produced across the analog output terminals.

 10. ERROR MESSAGES

Table 10.1 List of Error Messages

No.

Error Message

Description

1

ADJUST DATA ERROR

After adjusted, error of allowable range check

12

A/D CONVERT ERROR

A/D converter circuit error

13

DENSITY CONVERT ERROR

Density converter circuit error

14

TEMP 1 (PT) OVER

Temperature Pt input 1 overflow

15

TENP 1 (PT) UNDER

Temperature Pt input 1 underflow

16

TEMP 2 (PT) OVER

Temperature Pt input 2 overflow

17

TEMP 2 (PT) UNDER

Temperature Pt input 2 underflow

18

TEMP 1 (ANA) OVER

Analog temperature input 1 overflow

19

TEMP 1 (ANA) UNDER

Analog temperature input 1 underflow

20

TEMP 2 (ANA) OVER

Analog temperature input 2 overflow

21

TEMP 2 (ANA) UNDER

Analog temperature input 2 underflow

26

DENSITY OVER

Density input overflow (unused)

27

DENSITY UNDER

Density input underflow (unused)

28

4mA SCALER 1 UNDER

Analog output 1 underflow

29

20mA SCALER 1 OVER

Analog output 1 overflow

30

4mA SCALER 2 UNDER

Analog output 2 underflow

31

20mA SCALER 2 OVER

Analog output 2 overflow

NOTE: Depending on the model in service, some of the messages are not shown.

Summary of Contents for EL4131

Page 1: ...nowledge contained in this manual you can expect the best possible results over a long service life of this instrument Keep this manual for ready reference It is suggested that the instruction manuals...

Page 2: ...NTERNAL COMPONENTS AND FUNCTIONS 10 6 1 Front Panel 10 6 1 1 Display 10 6 1 2 On Screen Menu Items 10 6 1 3 Error Messages 10 6 1 4 Front Panel Keypad 11 7 CALCULATION FORMULAS 12 7 1 Implementation o...

Page 3: ...1 170 9 EL4501 175 10 DIP SWITCH CONFIGURATION 182 Model with communication capabilities COMMUNICATION MANUAL Ins No E 880CM E CONTENTS 1 COMMUNICATION CAPABILITY OVERVIEW 5 2 NETWORK CONFIGURATION 5...

Page 4: ...ire supply complete informa tion as to the product name model number product number ratings and other pertinent information 1 2 Transportation Precautions 1 In order to safeguard against damage during...

Page 5: ...the companion flowmeter is simple by keystrokes on the front panel keypad or by inserting an IC card into the slot 2 Built around a microprocessor the instrument carries out calculations entirely in d...

Page 6: ...ed to install instruments alongside each other dimensions in Fig 3 2 are suggested 3 Recommended mounting height is given in Fig 3 3 3 2 3 Installation 1 Front mount the instrument through the cutout...

Page 7: ...s are found on the back of the instrument Fig 4 1 4 3 Description of Terminal Blocks for External Connections Terminal blocks for external connections TB1 TB2 and TB3 are found on the back of the inst...

Page 8: ...4 to 20mADC 1 to 5VDC 5 Temp in b Pt100 3 wire 5 6 B 6 Pulse out 1 No Polarity Uncorrected total flow Static relay 7 A 7 8 4 to 20mADC 1 to 5VDC 8 Pulse out 2 No polarity Corrected total flow Static r...

Page 9: ...Same as output pulse unit m3 etc Display capacity 8 digits Corrected total flow Same as output pulse unit m3 C etc Uncorr instant flowrate m3 h etc Corr instant flowrate m3 h C etc Temperature To the...

Page 10: ...nd corrected for temperature is indicated 5 Temperature TEMPERATURE The temperature value currently fed to the instrument is indicated If a problem arises in the course of temperature conversion such...

Page 11: ...ion keys Four round keys Mainly used for function selection such as activating the input conditions selected and recon figuration Shift keys Four triangular keys Used for moving the cursor moving betw...

Page 12: ...ont panel keys and the frequency pulse information flowrate 7 2 Implementation of Temperature Correction Corrected total flow Q2 Q1 K K a b t0 t c t0 t 2 where Q2 Total flow corrected for temperature...

Page 13: ...ems of the total mass flow and total calorific flow and make sure that incoming pulses are being accumulated Also verify that pulse output and analog output are properly generated 5 Remove the simulat...

Page 14: ...ure correction factor 1 Temperature reading correct 1 A fault in internal assembly While fluid is allowed to flow Fails to count pulses fails to produce a pulse output 1 Input signal line correctly wi...

Page 15: ...e range check 12 A D CONVERT ERROR A D converter circuit error 13 DENSITY CONVERT ERROR Density converter circuit error 14 TEMP 1 PT OVER Temperature Pt input 1 overflow 15 TENP 1 PT UNDER Temperature...

Page 16: ...D Converter Press Input Ckt analog Temp Input Ckt analog Temp Input Ckt Pt Analog Output Circuit Pulse Output Circuit 1 Pulse Output Circuit 2 Pulse Output Circuit 3 Multi plexer Isolation Test Pins...

Page 17: ...e N Non Output item 2 Pulse output 2 B 1 Before correction pulse width 1ms B 5 Before correction pulse width 50ms B 9 Before correction special pulse width C 1 After correction pulse width 1ms C 5 Aft...

Page 18: ...am Japanese D W E Wiring diagram English D A J Compression coef cient calculation process Japanese D A E Compression coef cient calculation process English D T J Inspection procedure Japanese D T E In...

Reviews: