36 | Ottobock
ScoutCrawler
Fig. 19
Direction of the charging cable
7.4.5.3 Charging after Deep Discharge
In case of deep discharge of the battery pack, proceed as
follows:
1) Turn off the control unit on the product.
2) Remove the battery case cover (see Fig. 16, item 1).
3) Loosen the battery support strap (see Fig. 16, item 2).
4) Pull the release lever of the connection holder up (see
Fig. 17, item 1).
5) Loosen the connection holder by rocking it slightly and
remove it from the battery pack.
NOTICE! Do not use force to loosen the battery
pack connections.
6) Reinstall the connection holder on the battery pack.
NOTICE! Do not use force to connect the battery
pack connections.
→
The battery pack can be charged as described in
chapter 7.4.5.1 "Charging via the Control Panel".
7.5 Disassembly/transport
7.5.1 Safety Instructions
WARNING
Improper transportation in aircraft
Risk of injury due to fire or explosion, battery damage,
short circuit
►
Follow the IATA (International Air Transport Associ
ation) rules when transporting the product in an air
craft. To this end, the batteries must always be
removed and the battery connectors must be insulated
so that they are short-circuit-proof prior to check-in of
the product as luggage. For more information, please
visit www.iata.org.
►
Ottobock recommends contacting the airline directly
before every flight to obtain information regarding spe
cial transport regulations.
INFORMATION
Lithium-ion batteries are classified as hazardous goods.
Therefore the battery pack should either be installed in the
Usage