Remote Control
Genium - Bionic Prosthetic System
56 | Ottobock
Remote Control
Genium - Bionic Prosthetic System
56 | Ottobock
INFORMATION
The prosthetist's setting is usually found in the middle of the scale. After making changes, the setting established
by the prosthetist can be restored by moving the slider to the middle.
INFORMATION
The knee joint settings should be optimised by the prosthetist using X-Soft. The remote control is not intended for
use by the prosthetist in order to set up the knee joint. The day-to-day behaviour of the prosthesis can be changed
to a certain degree when the patient uses the remote control (e.g. as the user is becoming accustomed to the pros-
thesis). At the next appointment, the prosthetist can trace the changes in X-Soft.
The tables below show the adjustment parameters for the individual MyModes along with a brief description, the
adjustment range of the remote control around the value selected by the prosthetist and the absolute value range.
Basic mode:
X-Soft
Remote control Absolute range
Remote
control
adjustment range
Description
Stance phase
flexion resistance
Resistance
120-180
±10
Resistance against flexion motion
e.g. when walking down stairs or
when sitting down on a chair.
Swing flexion angle Angle
55°-70°
±3°
Maximum flexion angle in the swing
phase.
Pitch
Pitch
1000Hz-4000Hz
1000Hz-4000Hz
Frequency (pitch) of the beeper for
confirmation signals.
Volume
0-4
0-4
Volume of the beeper for confirma-
tion signals.
MyModes except knee braking mode:
X-Soft
Remote control Absolute range
Remote
control
adjustment range
Description
Increase
Increase
0-100
±10
Value for speed at which flexion re-
sistance increases as the knee ang-
le increases.
Pitch
Pitch
1000Hz-4000Hz
1000Hz-4000Hz
Frequency (pitch) of the beeper for
confirmation signals.
Volume
Volume
0-4
0-4
Volume of the beeper for confirmati-
on signals.
Knee braking mode:
X-Soft
Remote control Absolute range
Remote
control
adjustment range
Description
Basis
Basis
0-180
±20
Basic resistance for the knee joint
without load
Pitch
Pitch
1000Hz-4000Hz
1000Hz-4000Hz
Frequency (pitch) of the beeper for
confirmation signals.
Volume
Volume
0-4
0-4
Volume of the beeper for confirmati-
on signals.
7.7 Deep sleep mode
INFORMATION
Advise your patient of the information in this section.
The remote control can be used to place the knee joint in a deep sleep mode, in which power consumption is mini-
mised. The knee joint offers no functionality in this mode.
It can be awakened from deep sleep mode with the remote control or by connecting the charger.