2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative
humidity 20 % to 90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range:
–10 °C to +45 °C
Moisture:
relative humidity: 20% to 90%, non-condensing
Unacceptable environmental conditions
Chemicals/liquids:
fresh water, salt water, perspiration, urine,
acids, soapsuds, chlorine water
Solids:
dust, sand, highly hygroscopic particles (e. g. talcum)
2.4 Service life
This product was tested by the manufacturer with 3 million load
cycles. Depending on the user’s activity level, this corresponds to a
maximum service life of 5 years.
3 Safety
3.1 Explanation of warning symbols
CAUTION
Warning regarding possible risks of accident or
injury.
3.2 General safety instructions
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
►
Use the product according to the specified area of application
(see Page 6).
7
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
►
Only combine the product with prosthetic components that are
approved for that purpose.
►
Based on the instructions for use of the prosthetic components,
verify that they may be combined with each other.
CAUTION
Use under unallowable environmental conditions
Risk of injury due to damage to the product
►
Do not expose the product to unallowable environmental condi
tions.
►
If the product has been exposed to unallowable environmental
conditions, check it for damage.
►
If damage is apparent or in case of doubt, do not continue using
the product.
►
Take suitable measures if required (e.g. cleaning, repair,
replacement, inspection by the manufacturer or a specialist
workshop etc.).
CAUTION
Exceeding the service life and reuse on another patient
Risk of injury due to loss of functionality as well as damage to the
product
►
Ensure that the approved service life is not exceeded.
►
Only use the product on a single patient.
CAUTION
Mechanical damage to the product
Risk of injury due to change in or loss of functionality
►
Use caution when working with the product.
Summary of Contents for 2R10
Page 2: ...1 2 3 4 5 1 2 2...
Page 65: ...65 4 1 1 2 2S88 1 1 1 3 2 2D5...
Page 72: ...72 4 1 1 2 2S88 1 1 1 3 2 2D5...
Page 75: ...2 2 1 2 2 78 2 3 20 C 60 C 20 90 10 C 45 C 20 90 2 4 3 5 75 3 3 1 3 2 75...
Page 76: ...76 4 1 1 2 2S88 1 1 1 3 2 2D5 5...
Page 79: ...2 4 300 5 3 3 1 3 2 78 79 1...
Page 82: ...2 2 84 2 3 20 C 60 C 20 90 10 C 45 C 20 90 2 4 3 5 3 3 1 3 2 82 82...
Page 85: ...2 4 3 5 3 3 1 3 2 84 85 4 1 1...