CHAPTER SEVEN
OAE test procedure
66
OAE test procedure
Test results
The following table lists all possible test results with the associated result
graphic and gives an explanation of the circumstances under which each
result would be shown.
Test Result
Description
TEOAE Pass
The data collected has met the criteria set.
The optimum test setting will depend on your
application (see
11.3
Test settings
)
Note:
The following results will only occur if a TEOAE Pass
is not obtained
No Valid OAE
The data collected has not met the set pass criteria
and the test conditions were acceptable
Noisy
The noisy data collected is three times greater than
the low noise data collected
Poor Probe Fit
Probe movement has been detected
Too Few Bands
Insufficient bands meet their pass criteria
Stopped Too
The test has been ended manually before the
Soon
required minimum amount of data has been collected