156 CHAPTER SIXTEEN
Probe care
Probe care
16.2.1
Disassembling the probe
First, unplug the probe from the instrument. Remove the tip and then the
lid (fig. 1). Remove the coupler tubes by pushing the end of the tubes
down onto a hard surface (fig. 2). Pull out the tubes by gripping them (fig.
3). Never remove them by pulling on the cable. Finally, pull the coupler
tubes away from the probe (fig. 4).
16.2.2
Reassembling the probe
Fit the new coupler tubes to the probe assembly (fig. 1). Fit the outer shell
(figs. 2 and 3), followed by the lid (fig. 4). Click the lid into place using firm
finger pressure only. Finally, fit a new tip (fig. 4).
16.2.3
Notes:
• Fit a new tip for each test.
• Check that the coupler tubes are not contaminated before
fitting the tip.
• If the coupler tubes are contaminated,
replace them
. We
recommend fitting new coupler tubes at regular intervals
(approx every 20-40 tests) as a preventive measure.
• Perform weekly probe QA tests (see chapter
14
Quality
checks
).