PRODUCT TROUBLESHOOTING
OI-7530 FAULT CODES
57
Problem
Reason
Solution
F8
There are two sensors with the same address
communicating to the monitor. This could be two
different Gen II sensor assemblies, or a Gen I and Gen
II sensor assembly. The monitor cannot tell if there are
two different Gen I units communicating to the same
device with the same address.
Check all units and make sure they are all using a different
addresses.
F9
Radio timeout. This means the monitor has not gotten a
communication from the faulting address for over X
minutes. X is equal to the radio timeout that is set in the
start up menu options. It defaults to 10 minutes.
Find the sensor assembly and see why it is not communicating.
This could be due to a dead battery, broken antenna, bad antenna
cable, no antenna, obstacle, weather, etc.
F15
This fault is no longer assigned. If “F15” is displayed on a sensor assembly, the firmware should be updated