oticon REM AutoFit Quick Manual Download Page 5

5

Step 2: Measure

Note that for simultaneous binaural measurements, Audioscan recommends 
that the client is a distance of 45–60 cm (18–24 in.) from the loudspeaker. 
All measurements use the signal type selected in step 1: Setup. Each 
sequence of measurements is preceded by a sound field equalization burst. 

In on-ear mode, the instruments will mute automatically prior to the sound 
field equalization burst. This ensures that, when the instruments are 
automatically unmuted, the subsequent measurements are accurately 
calibrated because the soundfield calibration was not affected by acoustic 
leakage. This method is suitable for both open and closed fittings, but it is 
important that your client remains still during each measurement sequence. 

For test box mode, the instruments are not muted during the sound field  
equalization burst as the venting is occluded upon attachment to the coupler. 
Open fittings cannot be verified in test box mode.

In the Measure step, there are two ways of performing verification 
measurements: Automatic and Manual.

Automatic: measures the current instrument output, automatically adjusts 
it to match the prescribed target, and then performs another measurement to 
confirm the target match.

a. 

The automatic measurement process is carried out at 65 dB by default 
and you can select additional levels (50 dB and 75 dB). Gain adjustments 
are based on the 65 dB measurement. 

b. 

If performing on-ear measurements, instruct your client to face the 
loudspeaker and to remain still during each sequence of measurements.

c. 

Click Start on the relevant side to measure one side. Click Both to 
measure both sides (binaurally). Binaural measurements take place 
simultaneously if using Genie 2 | 2019.2 and Verifit2 with software 
version 4.18 or later. Otherwise they take place sequentially.

d. 

When doing simultaneous binaural measurements, if the signal reaching 
left and right reference microphones is not equal, e.g. due to the client 
not facing the speaker, the measurement will be paused and Verifit2’s 
Binaural Sound Field Assist tool will show instructions on the Verifit2 
screen to help balance the signal reaching left and right.

e. 

The test signal will turn off automatically when the measurements are 
complete. If you need to stop the measurement prior to completion,  
click Stop.

f. 

If you are satisfied with the measurements, proceed to step 3: Finish.  
If you want to perform further fine-tuning, repeat the automatic 
measurement, or click the Manual tab to perform a manual measurement. 
Note that if you repeat the automatic process immediately, an initial 
measurement will not be done.