oticon REM AutoFit Quick Manual Download Page 3

3

Settings in other tools

The following controls are set automatically when REM AutoFit with VerifitLINK 
is open and restored to their original states afterwards:

Adaptation Manager

Step 3

Noise reduction

Current setting in OpenSound Navigator™

Feedback Analyzer

On

Directionality setting

Pinna Omni. For styles with a single 
microphone, directionality is not applicable.

Speech Rescue

Off

Step-by-step Guide

On the Fitting step in Genie 2, enter the REM tool from the More Tools menu 
in the left taskpane. Launch REM AutoFit with VerifitLINK by clicking the 
VerifitLINK button at the bottom of the REM tool.

Step 1: Setup

 

1. 

Select a Measurement method from the drop-down list:  
On ear or Test box 

2. 

Select a Fitting rationale. You can use all available fitting rationales  
on the Verifit, except Camfit.