REM
AutoFit with
Verifit
®
LINK
Quick Guide
Page 1: ...REM AutoFit with Verifit LINK Quick Guide...
Page 2: ...or Verifit2 4 18 or later is required for simultaneous binaural measurements You must have a network connection between the Verifit and your computer If you are already using the Verifit NOAH module y...
Page 3: ...ni For styles with a single microphone directionality is not applicable Speech Rescue Off Step by step Guide On the Fitting step in Genie 2 enter the REM tool from the More Tools menu in the left task...
Page 4: ...D tool from NOAH 5 Prepare for the measurements by following the instructions to the right of the screen These instructions are specific to the measurement method you have selected One of the instruct...
Page 5: ...m the target match a The automatic measurement process is carried out at 65 dB by default and you can select additional levels 50 dB and 75 dB Gain adjustments are based on the 65 dB measurement b If...
Page 6: ...on the relevant side to measure one side Click Both to measure both sides binaurally Binaural measurements take place simultaneously if using Genie 2 2019 2 and Verifit2 with software version 4 18 or...
Page 7: ...ccording to what you select Select Apply all gain adjustments to apply adjustments Select Discard all changes to discard adjustments The data in use when you entered REM AutoFit with VerifitLINK will...