cmd
type
min
max
default
unit
description
GFD
float
1.2
24
5
V
default fan voltage
GX
bool
S
R
S
external control stop/run
GT
float
— no parameter —
◦
C
device temperature (head)
GVS
word
— no parameter —
software version
GVN
word
— no parameter —
serial number
9.3.1
Status
C
ommand
s
cmd
type
min
max
default
unit
description
GS
word
— no parameter —
get status
The status bits have the following meanings:
0x0001
interlock OK
0x0004
driver supply OK
0x0400
LT
sensor OK
0x0008
driver temperature OK
0x0800
CT
sensor OK
0x0010
LTLU
not OK
0x0020
LTLL
not OK
0x2000
LTM
not OK
0x0040
CTLU
not OK
0x4000
LC
ON
0x0080
CTLL
not OK
0x8000
LC
error
9.3.2
Mode
C
ommands
cmd
type
min
max
default
unit
description
GM
word
— no parameter —
get mode
GMC
word
clear mode bit(s)
GMS
word
set mode bit(s)
GMT
word
toggle mode bit(s)
The mode bits have the following meanings:
0x0001
laser current ON
0x0100
first TEC (laser) ON
0x0002
input echo OFF
0x0200
second TEC (crystal) ON
0x0400
pilot laser ON
0x0008
binary mode
0x0800
laser current control (
LCC
) OFF
0x0010
laser voltage control OFF
0x1000
use external interface after startup
0x0020
LMDI
ON
0x2000
LMDX
OFF
33