AFTER DISASSEMBLY / ASSEMBLY
Check the following:
• Are the duckbill valves oriented in the correct direction
(Figure 4)
?
• Is the membrane fully inserted into the housing?
• Are the inlet and outlet ports tightened?
• Is the tube fully inserted on the inlet port? (Check both at the Unity
Socket valve and at the Unity Membrane housing)
FINISHING / LINER OPTIONS
Össur recommends that the Unity System is used with an Iceross Seal-In
liner.
caution
: Make sure the Flex-Foot sock does not interfere with the Unity
System as it could compromise the function of the pump.
THE TUBE
The tube is the most sensitive part of the Unity System and special
attention should be paid to it for potential functional issue that might
arise.
CAUTION:
Ensure that Unity tubing is not compressed or folded by the
application of cosmetic finishing.
Make sure the tube is NOT:
• Folded
(Figure 3)
.
• Pressed up against sharp objects.
• Punctured.
• Dangling during gait.
• Located laterally, as there is more likelihood of damage.
IN CASE THE UNITY SYSTEM IS NOT FUNCTIONING PROPERLY
Check the following:
• Is there a leakage in the socket?
• Is the tube punctured?
• Is the membrane punctured or torn?
• Has the membrane been pulled out of the membrane housing?
• Is there dust or grit inside the system?
• Is the tube fully inserted on the inlet port? (Check both at the Unity
Socket valve and at the Unity Membrane housing)
• Are the inlet and outlet ports tightened?
• Is the Unity Socket valve leaking?
• Is the sock inside the foot cover affecting the systems mechanical
movements?
MAINTENANCE
The following components of the Unity System might need cleaning or
replacing during the lifetime of the product. This is to be considered
normal wear and is not covered by the Limited Product Warranty.
• Membrane
• Duckbill valves
• Tube
It is recommended to have the product inspected every 6 months by a
clinical specialist.
Clean the components with mild soap and rinse with water, isopropyl
alcohol or Iceross Clean&Simple spray.
DO NOT use solvents stronger than isopropyl alcohol.
Summary of Contents for UNITY
Page 1: ...Instructions for Use UNITY FOR ASSURE BALANCE FOOT J...
Page 4: ...4 0 5 1 5 mm 0 5 1 5 mm 0 5 mm A B C 3Nm A B C D A B C 5 C A B 6 7...
Page 56: ...56 www ossur com Unity System 24 12 Iceross...
Page 60: ...60 Copolymer Gel Iceross...
Page 64: ...64 Unity 6 IcerossClean Simple ssur www ossur com Unity 24 12 Iceross...