20
CATEGORY SELECTION CHART
Please refer to the selection charts below to determine the appropriate stiffness required according to Össur
recommendations.
LIABILITY
The manufacturer recommends using the device only under the specified conditions and for the intended
purposes. The device must be maintained according to the instructions for use. The manufacturer is not
liable for damage caused by component combinations that were not authorized by the manufacturer.
COMPLIANCE
This component has been tested according to ISO 10328 standard to two million load cycles. Depending on
the amputee’s activity this corresponds to a duration of use of two to three years. We recommend carrying
out regular yearly safety checks
Recommended separation distances between portable and mobile Rf communications equipment and the device.
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d=1,2√P
80 MHz to 800 MHz
d=1,2√P
800 MHz to 2,5 GHz
d=2,3√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
USeR infoRmaTion kg
45-52
53-59
60-68
69-77
78-88
89-100
101-116
117-125
USeR infoRmaTion lbS
99-115
116-130
131-150
151-170
171-194
195-221
222-256
257-275
Low Impact Level
1
1
2
3
4
5
6
7
Moderate Impact Level
1
2
3
4
5
6
7
8
Summary of Contents for PROPRIO FOOT