Four-Point Probe System
User Manual
10
Ossila Limited © 2022
6.
Specifications
The Four-Point Probe System measurement specifications are shown in
and the physical specifications are shown in
Table 6.1.
Four-Point Probe System measurement specifications.
Voltage range
±
100 μV to
±10 V
Current range
±10 nA to ±200 mA
Sheet resistance range
100
mΩ/
square to 10
MΩ/
square
Table 6.2.
Accuracy and precision of the Four-Point Probe System at each resistance range. Accuracy is the
maximum deviation from the true value. Precision is the maximum deviation between identical
measurements (useful for comparative measurements).
Sheet Resistance
Accuracy
Precision
Measured at Range
100 mΩ/
square
±8%
±3%
200 mA
1 Ω/
square
±2%
±0.5%
200 mA
10 Ω/
square
±1%
±0.5%
200 mA
100 Ω/
square
±1%
±0.05%
20 mA
1 kΩ/
square
±1%
±0.03%
20 mA
10 kΩ/
square
±1%
±0.02%
2 mA
100 kΩ/
square
±2%
±0.05%
200 μA
1
MΩ/
square
±8%
±0.5%
20 μA
10 MΩ/
square
±30%
±5%
20 μA
Table 6.3.
Four-Point Probe System physical specifications.
Probe Spacing
1.27 mm
Rectangular Sample
Size Range
Long Edge Minimum: 4 mm
Short Edge Maximum: 60 mm
Circular Sample Size
Range (Diameter)
4 mm to 76.2 mm
Maximum Sample
Thickness
10 mm
Overall Dimensions
Width: 145 mm
Height: 150 mm
Depth: 240 mm