Four-Point Probe System
User Manual
16
Ossila Limited © 2022
(III)
Current Range
•
Select whether the appropriate range should be automatically determined or whether a
specific range of currents is used.
I.
This defines the upper limit and accuracy of current measurements that can be
performed by the unit.
II.
Autorange will enable the measurement to automatically switch ranges as needed.
III.
The maximum current values for each range are shown in the range selection box.
(IV)
Polarity
•
Sets whether the measurement polarity is positive or negative.
(V)
Voltage Limit
•
Sets the maximum voltage the measurement can apply to the sample in volts.
I.
If the voltage limit is exceeded the measurement will stop.
(VI)
Current Limit
•
Sets the maximum current the measurement can apply to the sample in mA.
I.
If the current limit is exceeded the measurement will stop.
(VII)
Reset to Default
•
Clicking this button will reset all advanced settings to their default values.