Orthofix PILLAR SA PTC Instructions For Use Manual Download Page 3

3

AW-39-9901  Rev.AA

PK-1602-PL-US © Orthofix, Inc.  7/2016

Provided Non-Sterile

Serial Number

Authorised 

Representative

CAPA Symbols  Chart revised

See Instructions for Use

 

Sterilized Using

Irradiation

 

Do Not Resterilize

 

Lot Number

Manufacturer

Use By Date

Federal (U.S.A.) law restricts this device to sale 

by or on the order of a physician

Single Use Only

Do Not Reuse

 

Catalogue Number

 

Manual Cleaning:

 

1.  Upon completing the preparation for the cleaning procedure, prepare the Vesphene

®

 IIse 

agent per the dilution recommended on the label directions (1 ounce per gallon), or 1 mL 

of Vesphene

®

 IIse to each 128 mL of potable tap water per the manufacturer’s Directions 

for Use label. 

2.  Bathe instruments in prepared room temperature solution.

3.  Manually agitate instruments in Vesphene

®

 IIse solution for 15 minutes.

4.  If visible soil is noted, scrub instruments with a soft plastic bristle brush and use the 

brush or a pipe cleaner long enough to reach the entire length of any interior lumen(s) to 

remove the soil.

5.  Rinse the instruments in USP <1231> purified water for 1.5 minutes.

6.  Hang dry the device.

7.  Inspect the instruments for visible soil.

8.  If visible soil is noted, repeat the steps listed above.

Automated Cleaning:

 

1.  Upon completing the preparation for the cleaning procedure, set up the washer/disinfector 

detergent dose at ½ ounce of Endozime AW Plus

®

 per gallon of water or according to the 

manufacturer recommendations.

2.  If visible soil is noted, scrub instruments with a soft plastic bristle brush and use the 

brush or a pipe cleaner long enough to reach the entire length of any interior lumen(s) to 

remove the soil.

3.  Place scrubbed instruments into the washer baskets.

4.  Orient instruments into the automated washer’s carriers as recommended by the washer 

manufacturer.

5.  The following automated cleaning cycle is recommended (minimum recommended times 

are provided for each stage):

a.  Pre-Wash 1: cold potable water, 2 minutes

b.  Enzyme/Detergent treatment:

1.  Spray with enzyme/detergent, 20 seconds

2.  Soak, 1 minute

3.  Rinse cold potable water, 15 seconds

4.  Rinse cold potable water, 15 seconds

c. Wash 

 65°C, 2 minutes using Endozime AW Plus

®

d.  Rinse 1: hot potable water, 15 seconds

e.  Rinse 2: hot potable water, 15 seconds

f.  Rinse 3: hot potable water, 15 seconds

g.  Rinse 4: hot potable water, 15 seconds

h.  Thermal rinse 

 93°C, 1 minute

i.  Heated USP <1231> Purified Water Rinse 1: re-circulating 10 seconds

j.  Heated USP <1231> Purified Water Rinse 2: non re-circulating 10 seconds

k.  Dry at 115°C, 7 minutes 

6.  Inspect the instruments for visible soil.

7.  If visible soil is noted, repeat the above listed steps until no visible soil is noted.

Note: 

 Certain cleaning solutions such as those containing caustic soda, formalin, glutaralde-

hyde, bleach, and/or other alkaline cleaners may damage instruments. These solutions should 

not be used.  

Note:

 Visually inspect instruments after cleaning and prior to each use. Discard or return to 

Orthofix any instruments that are broken, discolored, corroded, have cracked components, pits, 

gouges, or are otherwise found defective. Do not use defective instruments.

Sterilization:

The PILLAR SA PTC spacers are provided STERILE. They are sterilized using gamma irradiation 

sterilization. Do not re-sterilize. 

Sterilization in Orthofix Cases with Blue Wrap:

The PILLAR SA bone screws, plates and instruments are supplied NON-STERILE. Prior to use, all 

non-sterile instruments and implants should be placed in the appropriate Orthofix case which 

will be wrapped in a FDA cleared sterilization wrap and placed in the autoclave for sterilization 

by the hospital using one of the following recommended cycles:

 

Method: Steam  

or:  

Method: Steam 

 

Cycle: Gravity  

 

Cycle: Prevac 

 

Temperature: 270°F (132°C)  

 

Temperature: 270°F (132°C) 

 

Exposure time: 15 minutes 

 

Preconditioning: 4 Pulses

 

Drying Time: 30 minutes 

 

Exposure time: 4 minutes

 

Double wrapped 

 

Drying Time: 30 minutes

 

 

 

Double wrapped

Validation and routine monitoring should be performed per ANSI/AAMI ST79 

Comprehensive 

guide to steam sterilization and sterility assurance in health care facilities. Other cycles may be 

used as long as they comply with the above practices and provide a sterility assurance level of 

10

-6

.

Packaging:

Packages for each of the components should be intact upon receipt. If a consignment system 

is used, all sets should be carefully checked for completeness and all components should be 

carefully checked for damage prior to use. Damaged packages or products should not be used 

and should be returned to Orthofix. 

The PILLAR SA PTC instruments, bone screws, and plates are provided in a modular case specifi-

cally intended to contain and organize the system’s components. The system’s instruments are 

organized into trays within each modular case for easy retrieval during surgery. These trays 

also provide protection to the system components during shipping. Additionally, individual 

instruments and implants are provided in sealed poly bags with individual product labels.

PILLAR SA PTC spacers are provided STERILE. Do not use if the package is opened or damaged 

or if the expiration date has passed.

Product Complaints:

Any Health Care Professional (e.g., customer or user of this system of products) who has any 

complaints or who has experienced any dissatisfaction with the product quality, identity, durability, 

reliability, safety, effectiveness and/or performance, should notify Orthofix Inc., 3451 Plano 

Parkway, Lewisville, TX 75056, USA, by telephone at 1-214-937-3199 or 1-888-298-5700 or by 

e-mail at [email protected]

Further Information:

A recommended Operative Technique for the use of this system is available upon request from 

Orthofix at the phone numbers provided above. 

Latex Information:

The implants, instruments and/or packaging material for the PILLAR SA PTC Spacer System are 

not formulated with and do not contain natural rubber. The term “natural rubber” includes 

natural rubber latex, dry natural rubber, and synthetic latex or synthetic rubber that contains 

natural rubber in its formulation.

 

Caution:

 Federal law (USA) restricts these devices to sale by or on the order of a physician.

Summary of Contents for PILLAR SA PTC

Page 1: ...tem Click directory below for desired language INSTRUCTIONS FOR USE Important Information Please Read Prior to Use Orthofix Inc 3451 Plano Parkway Lewisville Texas 75056 9453 U S A 1 214 937 3199 1 88...

Page 2: ...r motor function impotence perma nent pain and or deformity In rare instances some complications may be fatal Warnings and Precautions The surgeon should be aware of the following when using implants...

Page 3: ...ents that are broken discolored corroded have cracked components pits gouges or are otherwise found defective Do not use defective instruments Sterilization The PILLAR SA PTC spacers are provided STER...

Page 4: ...vios perif ricos mbolos pulmonares p rdida de la funci n sensorial o motora impotencia dolor permanente o deformidad En casos raros algunas complicaciones podr an ser fatales Advertencias y precaucion...

Page 5: ...luciones de limpieza alcalinas podr an da ar los instrumentos Estas soluciones no deber n utilizarse Nota Examine visualmente los instrumentos despu s de la limpieza y antes de cada uso Deseche o devu...

Page 6: ...ou difformit Dans de rares cas certaines complications peuvent tre mortelles Avertissements et pr cautions Le chirurgien doit prendre connaissance des avertissements suivants lors de l utilisation d...

Page 7: ...rald hyde de l eau de Javel et ou d autres nettoyants alcalins peuvent endommager les instruments Ces solutions ne doivent pas tre utilis es Remarque Inspecter visuellement les instruments apr s le ne...

Page 8: ...e o deformit In casi rari alcune complicanze possono essere fatali Avvertenze e precauzioni Nell uso degli impianti il chirurgo deve essere consapevole dei fattori elencati di seguito 1 La scelta cor...

Page 9: ...tica formalina glutaraldeide candeggina e o altri detergenti alcalini possono danneggiare gli strumenti Non usare queste soluzioni detergenti Nota esaminare visivamente gli strumenti dopo la pulizia e...

Page 10: ...Deformit ten Einige der Komplikationen k nnen in seltenen F llen letal sein Warn und Vorsichtshinweise Bei der Verwendung von Implantaten muss der Chirurg folgendes beachten 1 Die richtige Auswahl des...

Page 11: ...inigungsmittel enthalten k nnen Instrumente besch digen Diese L sungen d rfen nicht verwendet werden Hinweis Nach der Reinigung und vor jedem Gebrauch eine Sichtpr fung der Instrumente durchf hren Ins...

Page 12: ...r g rmesi pulmoner emboli duyu ve veya motor i levi kayb impotans kal c a r ve veya deformite Nadir durumlarda baz komplikasyonlar l mc l olabilir Uyar lar ve nlemler Cerrah implantlar kullan rken a a...

Page 13: ...eya di er alkali temizleyiciler i erenler gibi baz temizleme sol syonlar aletlere zarar verebilir Bu sol syonlar kullan lmamal d r Not Temizlemeden sonra ve her kullan m ncesinde aletleri g zle incele...

Page 14: ...sensorial e ou motora impot ncia dor permanente e ou deforma o Em casos raros algumas complica es podem ser fatais Avisos e precau es O cirurgi o deve estar ciente do seguinte ao utilizar implantes 1...

Page 15: ...de limpeza alcalinos podem danificar os instrumentos N o deve utilizar estas solu es Nota Inspeccione visualmente os instrumentos ap s a limpeza e antes de cada utiliza o Elimine ou devolva Orthofix...

Page 16: ...orische functie impotentie permanente pijn en of vervorming In zeldzame gevallen kunnen sommige complicaties dodelijk zijn Waarschuwingen en voorzorgsmaatregelen De chirurg dient zich bij het gebruik...

Page 17: ...igingsmiddelen kunnen de instrumenten beschadigen Deze oplossingen mogen niet worden gebruikt Opmerking Controleer de instrumenten op het oog na reiniging en v r elk gebruik Instrumenten die gebroken...

Page 18: ...4 5 1 2 3 4 5 6 7 8 9 1 2 3 PILLAR SA PTC 4 PILLAR PTC 5 PILLAR SA 6 PILLAR SA PTC 7 PILLAR SA PEEK 8 9 10 11 PILLAR SA PTC MR MR PILLAR SA PTC PILLAR SA PILLAR SA PEEK PILLAR SA PTC Orthofix Inc 345...

Page 19: ...P 1231 1 10 j USP 1231 2 10 k 115 C 7 6 7 Orthofix PILLAR SA PTC Orthofix PILLAR SA Orthofix FDA 132 C 270 F 132 C 270 F 15 4 30 4 30 ANSI AAMI ST79 Comprehensive guide to steam sterilization and ster...

Reviews: