13
Options Menu
Set/check the angle of the cup by
navigating impactor.
NOTE:
Title changes to “Check Cup
Angle” after cup angle has been
“Set”.
Measure change in leg length and offset
from its pre-operative state.
M1
Selection Option
Set Cup Angle
Check Leg Length
Verify Homepoint
Repeat Registrations
End Procedure
Re-register Homepoint. The change in position
of probe tip between original and current
position is displayed.
Repeat pelvic landmark registrations.
Use to select other hip for a bi-lateral procedure.
The unit can also be powered off at this point by
pressing and holding left and up arrow buttons
until unit powers off.
Prepare Impactor and Shell
Select the appropriate adapter as listed on page 7 of this surgical technique.
Straight Impactor:
Snap the adapter into the impactor.
Thread the adapter into the acetabular shell.
The release lever can be used to release the adapter from the impactor.
Offset Impactor:
Snap the adapter into the impactor.
Thread the adapter into the acetabular shell.
Pull on the finger grip shown to release the adapter from the
impactor.
Following full seating of the shell, remove the impactor from the
implant adapter.
Use the socket driver to remove the adapter from the implant.
NOTE:
Align the screw holes of shell as desired; Adapter can be assembled in multiple angles relative to the impactor.
Vertical orientation of the Reference Sensor post is not required for accuracy; it is recommended for ease of use when
transferring the Reference Sensor to the impactor for navigation.
If none of available adapters fits on selected implant, impactor cannot be navigated.
WARNING: Prior to impacting, verify shell is fully seated on shoulder of impactor and that threads do not protrude beyond
outer face of shell.
Release Lever
Pull to release adapter