6
6
Place on left Side
Calibration Steps
TIPS FOR CALIBRATION STEPS:
Navigation unit must be maintained in steady position, or calibration will not be accepted. A red hand will flash
on screen if unit is not steady.
Level circle must be green, but ball does not need to be in exact center of circle.
Do not move to next position too quickly. Wait for double beep, then place in next position as shown on screen.
Displayed numbers are the percent errors from the nominal jig angles on each axis, and should be 4.0 or less.
If navigation unit does not proceed to next screen after several seconds, check that navigation unit and reference
sensor are correctly attached to the bracket and press left button three times to repeat from first calibration step
(Step 4).
STEP 4:
Angle unit away from you so that pelvic jig rests on table
with navigation unit screen facing up.
STEP 5:
Angle unit toward you so that reference sensor rests on
table with navigation unit screen facing you.
STEP 6:
Rest navigation unit on left side on table.
STEP 7:
Angle unit away from you so that pelvic jig rests on table
with navigation unit screen facing up.
STEP 8:
Attach reference sensor to coupler on probe. Verify that
numbers are displayed on the screen by sliding the probe to
different positions.
4
Angle Forward
Cal 1
Cal 2
Cal 3
5
Angle Back
✔
Cal 1
Cal 2
Cal 3
8
Attach Sensor
Verify probe values are
displayed
102