3.3.

 

Positioning the ICS and ICS Stand with Inadequate Shield Clearance 

CAUTION: This procedure requires careful handling and alignment to prevent possible damage to the detector. It is 

recommended to keep the protective cover on the endcap when installing. Additionally, this installation is NOT 

recommended for a single person, and multiple individuals should be present to assist. 

1)  With the ICS Stand fully lowered, carefully roll the ICS assembly to the front of the lead shield. 
  

 

 

2)  Carefully lift and tilt the ICS stand and ICS detector towards off the floor and towards its back as if to 

lay it on its back. 

 
3)  Now rotate the entire assembly underneath the shield while simultaneously aligning the detector 

endcap into the access hole in the bottom of the shield. 

 
4)  Once centered tilt the assembly upright so that the detector endcap is positioned through the bottom 

shield opening and the ICS assembly is resting on the floor. 

 

 

 

Summary of Contents for ICS Stand II

Page 1: ...ORTEC Part No 1108265 Manual Revision B ORT X4548 ICS Stand II User Manual...

Page 2: ......

Page 3: ...n 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause unnecessary delays in getting the u...

Page 4: ...icates a hazard that could result in death or serious bodily harm if the safety instruction is not observed WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not...

Page 5: ...ACKING THE ICS STAND 5 1 5 THE ICS STAND 6 2 ICS INSTALLATION ONTO THE ICS STAND 8 2 1 THE ICS AND ICS STAND ASSEMBLY 8 3 INSTALLATION UNDER A TYPICAL 4 TOP LOADING LEAD SHIELD 12 3 1 Preparing the le...

Page 6: ...CIFICATIONS1 1 1 1 Performance Maintenance The ICS Stand doesn t require periodic maintenance Consult the detector and ICS manuals for regular maintenance of the HPGe detector and the cryocooler 1 1 2...

Page 7: ...nents The ICS Stand main assembly User manual and documentation package Pneumatic air pump for units shipped after November 2015 1 4 UPACKING THE ICS STAND 1 Unpack the ICS Stand and make sure the sta...

Page 8: ...1108265B ICS Stand 2 6 1 5 THE ICS STAND Figure 1 illustrates the main features of the ICS Stand shown without the ICS and detector...

Page 9: ...ICS Stand 2 1108265B 7 Figure 1 ICS Stand 2 fully compressed Mount panel Air piston Schrader air valve Directional selection lever Adjustable feet 4 Plug support plate...

Page 10: ...ncluded One pneumatic air pump with a Schrader valve such as a bicycle pump included 2 1 THE ICS AND ICS STAND ASSEMBLY 1 Remove the plug support plate by lifting the plate off the main bracket no too...

Page 11: ...ICS Stand 2 1108265B 9 3 Flip the directional valve to the up position 4 Use the air pump to raise the platform about 6 inches 150 mm...

Page 12: ...1108265B ICS Stand 2 10 5 Place ICS on the platform ICS P4 shown...

Page 13: ...late No tools are necessary The plate slides into position Installation of the plate is meant to be snug If a secondary plug is not used it is not necessary to reinstall the plug plate NOTE The ICS an...

Page 14: ...y accommodate the size of the detector capsule and endcap 3 Verify the height of the shield bottom and or lip of the shield stand and whether or not the ICS Stand in the fully lowered position will ha...

Page 15: ...he detector capsule with the center of the access hole in the shield bottom 3 Once centered attach the pneumatic air pump move the directional selection lever on the front of the ICS Stand to the UP p...

Page 16: ...per position is reached Disconnect the pneumatic air pump and move the directional selection lever to the LOCK position CAUTION The shield should not electrically contact the cryostat Plastic sheeting...

Page 17: ...should be present to assist 1 With the ICS Stand fully lowered carefully roll the ICS assembly to the front of the lead shield 2 Carefully lift and tilt the ICS stand and ICS detector towards off the...

Page 18: ...the detector capsule with the center of the access hole in the shield bottom 6 Once centered attach the pneumatic air pump move the directional selection lever on the front of the ICS Stand to the UP...

Page 19: ...per position is reached Disconnect the pneumatic air pump and move the directional selection lever to the LOCK position CAUTION The shield should not electrically contact the cryostat Plastic sheeting...

Page 20: ...ND 2 IN POSITION This is done after the ICS is installed on the stand and the ICS Stand II is in its desired location in the shield Using the hex wrench provided torque the lock nut to a maximum of 13...

Reviews: